Ocular Spot Fluorometer Equipped With a Lock-In Amplifier for Measurement of Aqueous Flare
Rachapalle Reddi Sudhir1, P Pavani Murthy1, Sirisha Tadepalli2
1Department of Cornea and Refractive Surgery, Sankara Nethralaya, Chennai, India.
Purpose:
To evaluate a custom-made ocular fluorometer for detection of intensity of light scatter (ILS) from the anterior chamber (A/C) as an objective measure of aqueous flare.
Methods:
The fluorometer, equipped with a lock-in amplifier, was employed in the scatter mode to detect ILS from A/C. Measurements were performed with two illumination slit widths of 0.5 and 0.25 mm. The axial resolution at these slit widths were 80 and 200 μm, respectively. Healthy and pseudophakic eyes, with grade 0 Standardization of Uveitis Nomenclature (SUN) score, were employed as control subjects. ILS was also recorded in a cohort of patients who had undergone phacoemulsification and showed grades 1+ or 2+ on postoperative days 1 and 4.
Results:
The inter- and intraobserver variabilities in the measurement of ILS were not significant. In cataract patients, ILS was significantly higher on postoperative day 1 relative to healthy eyes. By day 4, ILS decreased significantly and was only marginally different from ILS in quiet pseudophakic eyes or healthy eyes. Eyes with higher SUN scores showed proportionately increased ILS. The receiver-operator characteristic analysis indicated no advantage in using the smaller slit width in discriminating ILS at different SUN scores although it provided higher axial resolution.
Conclusions:
The lock-in-based spot fluorometer is reliable for measurement of ILS with high precision and accuracy.The measured ILS correlates linearly with SUN scores and can be used to provide a higher granularity for recording aqueous flare.
Translational Relevance:
The instrument can be used in the clinical management of uveitis and drug development toward uveitis.
Related Concept Videos
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Chemical Reactions in Aqueous Solutions
Operational Amplifiers
MOSFET Amplifiers
BJT Amplifiers
In BJT amplifier configurations, particularly in common-emitter setups, the transistor's role...