[Quantitative volumetric analysis of the lower extremity: validation against established tape measurement and water
Konstantin Christoph Koban1, Virginia Titze1, Lucas Etzel1
1Klinikum der Ludwig-Maximilians Universität, München Handchirurgie, Plastische Chirurgie, Äshetische Chirurgie.
Background:
Tape measures continue to be used for the diagnosis and evaluation of lipoedema and lymphoedema in the clinical routine. Extremity volumes are calculated based on standardised circumferential measurements. Other volume measurements such as water displacement are not applied on a regular basis. This study aimed to evaluate a 3D scanner as an alternative and reproducible tool to objectively measure the volumes of the lower extremity.
Patients, Materials And Method:
We used a commercially available 3D scanner, the Artec Eva®, to perform 3D scans of the lower extremities on 20 subjects and calculated the volume using the appropriate software. In addition, limb volume was determined with standardised methods using the circumferential method (cone method and disc method) and the water displacement technique. The results and the time taken to perform the three measurement methods were documented and statistically evaluated.
Results:
There were no significant deviations between the results from 3D volumetry and water displacement (p > 0.05). The cone method significantly overestimated the volume measured by water displacement and 3D volumetry. The disc method revealed no statistically significant differences, but clinically relevant differences with a high variance in the 95 % confidence interval. All methods demonstrated high positive correlations with each other. Water displacement was the most timeconsuming method.
Conclusion:
Our findings obtained from the examination of 40 lower extremities show that 3D scans and software-based volumetric calculations can achieve objective and reproducible measurements in a relatively short time. The deviation from the gold-standard method of water displacement is small. Compared with the cone and disc method, however, we observed clinically relevant deviations with a high variance. We therefore recommend the 3D scan procedure for the objective documentation of conservative and surgical treatments for lipoedema or lymphoedema and have incorporated it into our clinical routine.
More Related Videos
08:25Combining Volumetric Capnography And Barometric Plethysmography To Measure The Lung Structure-function Relationship
Published on: January 8, 2019
08:45Kinematic Analysis Using 3D Motion Capture of Drinking Task in People With and Without Upper-extremity Impairments
Published on: March 28, 2018
Related Concept Videos
Distance Measurements by Taping
Quantitative Analysis
In quantitative analysis, two key measurements are made: the sample quantity and a property proportional to the amount of the analyte (the substance being analyzed). This forms the basis of the...
Reliability and Validity
Errors in Taping
Displacement Current
Position and Displacement