Use of a Repositionable and Fully Retrievable Aortic Valve in Routine Clinical Practice: The RESPOND Study and
Nicolas M Van Mieghem1, Jochen Wöhrle2, David Hildick-Smith3
1Interventional Cardiology, Thoraxcenter, Erasmus Medical Center, Rotterdam, the Netherlands.
Objectives:
The authors sought to evaluate 1-year clinical outcomes with the Lotus valve (Boston Scientific, Marlborough, Massachusetts) in a large international, multicenter prospective registry including patients eligible for transcatheter aortic valve replacement (TAVR) based on heart team consensus.
Background:
TAVR is a safe and effective treatment for severe aortic valve stenosis; however, limited data are available on TAVR with the repositionable and fully retrievable Lotus valve in unrestricted contemporary clinical practice.
Methods:
The RESPOND (Repositionable Lotus Valve System-Post-Market Evaluation of Real World Clinical Outcomes) study enrolled 1,014 patients; 996 patients were implanted with the Lotus valve (mean age 80.8 years, 50.8% female, mean STS score 6.0 ± 6.9%). The primary endpoint was all-cause mortality in the intent-to-treat population at 30 days and 1 year. An Extension cohort of 50 patients was treated with the Lotus valve with Depth Guard including a modified delivery system. Mortality and stroke were independently adjudicated. An independent core laboratory assessed echocardiographic data.
Results:
One-year clinical follow-up was available for 99.9% of Lotus valve-treated patients. At 1 year, the all-cause mortality rate was 11.7% and 4.1% of patients had experienced a disabling stroke. The permanent pacemaker implantation rate was 32% (37% among pacemaker-naive patients). Echocardiographic data at 1 year were available for core laboratory assessment in 62.6% of patients. Paravalvular leak was absent or trace in 94.5%, mild in 5.1%, and moderate in 0.4% of patients. Data from the Extension cohort confirmed good clinical outcomes at 30 days with an 18% permanent pacemaker rate (20% among pacemaker-naive patients).
Conclusions:
One-year outcomes from the RESPOND study confirm the safety and efficacy of the Lotus valve when used in routine clinical practice. (Repositionable Lotus Valve System-Post-Market Evaluation of Real World Clinical Outcomes [RESPOND]; NCT02031302).
Related Concept Videos
Aortic Regurgitation II: Clinical Features and Diagnostic Tests
Retrieval
Recall involves accessing information without cues, such as during an essay test, where individuals must retrieve facts and concepts from memory unaided. Another example is remembering the name of a colleague...
ER Retrieval Pathway
The ER uses many checkpoints to prevent the entry of incorrectly folded or a resident protein as cargo onto a transport vesicle. These mechanisms...
Aneurysm II: Clinical Manifestations and Diagnostic Studies
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Hypertension III: Clinical Manifestations and Diagnostic Studies