An electrochemical aptasensor for staphylococcal enterotoxin B detection based on reduced graphene oxide and gold
Somayeh Mousavi Nodoushan1, Navid Nasirizadeh2, Jafar Amani1
1Applied Microbiology Research Center, Systems Biology and Poisonings Institute, Baqiyatallah University of Medical Sciences, Tehran 14359-44711, Iran.
Abstract:
Detection of staphylococcal enterotoxin B (SEB) as a bacterial toxin causing severe food poisoning is of great importance. Herein, we developed an electrochemical aptasensor for SEB detection using a screen printed electrode modified with reduced graphene oxide (rGO) and gold nano-urchins (AuNUs). Afterward, the single-stranded DNA probe was attached to the surface of AuNUs on the modified electrode and then the specific aptamer was attached to the probe. In the presence of SEB molecules, the aptamer detached from the electrode surface and after applying the electrochemical signal generator, hematoxylin and the peak current of differential pulse voltammetry (DPV) were recorded. Due to the intercalation mechanism of hematoxylin-DNA interaction, the detachment of aptamer from electrode surface decreased the DPV peak current and the calibration graph (peak current vs SEB concentration) can be used for quantification of SEB. The cyclic voltammetry (CV) and electrochemical impedance spectroscopy (EIS) and also field emission scanning electron microscope imaging were used for electrode characterization. Selectivity experiments of the developed aptasensor showed a very distinct difference between SEB and other nonspecific molecules. A wide linear range from 5.0 to 500.0 fM was achieved and the detection limit was calculated as 0.21 fM. The performance of the aptasensor was checked in spiked food samples as simulated real samples and the results showed no significant difference compared to the synthetic samples. Results of selectivity and repeatability of the aptasensor were satisfactory. In addition, better recovery percentages and also lower standard deviation of aptasensor compared to a commercial ELISA kit of SEB detection proved the superior performance of the proposed aptasensor.
More Related Videos
Related Concept Videos
Redox Titration: Other Oxidizing and Reducing Agents
What is an Electrochemical Gradient?
The chemical gradient relies on differences in the abundance of a substance on the outside versus the inside of a cell and flows from areas of high to low ion concentration. In contrast, the electrical gradient revolves around an...
Oxidation Numbers
Oxidation-Reduction Reactions
Pyruvate Oxidation
First, the enzyme pyruvate dehydrogenase removes the carboxyl group from pyruvate and releases it as carbon dioxide. The stripped molecule is then oxidized and releases electrons, which are then picked up by NAD+...
Reducing Line Loss
With a step-up transformer at the source, the voltage is increased, thereby reducing the current in the transmission lines since power loss in...