Image Artifacts in Optical Coherence Tomography Angiography Among Patients With Multiple Sclerosis
Mustafa Iftikhar1, Sidra Zafar1, Natalia Gonzalez2
1a Wilmer Eye Institute , Johns Hopkins University School of Medicine , Baltimore , MD , USA.
Artifacts are common in optical coherence tomography angiography (OCT-A) imaging, particularly motion artifacts in multiple sclerosis (MS) patients. These artifacts can affect vessel density measurements, requiring careful review in future studies.
Area of Science:
- Ophthalmology
- Neurology
- Medical Imaging
Background:
- Optical coherence tomography angiography (OCT-A) is a key imaging technique for visualizing retinal vasculature.
- Understanding image quality and potential artifacts is crucial for accurate interpretation, especially in neuro-ophthalmic diseases like multiple sclerosis (MS).
Purpose of the Study:
- To evaluate the prevalence and types of artifacts in OCT-A images of multiple sclerosis (MS) patients and healthy controls.
- To identify clinical factors associated with these artifacts.
- To assess the impact of artifacts on quantitative vessel density measurements.
Main Methods:
- Prospective cross-sectional study involving 102 participants (MS patients and healthy controls).
- OCT-A imaging performed using a Spectralis® OCT-A prototype (Heidelberg, Germany).
- Analysis of 385 images to categorize artifact types, frequency, and assess impact on vessel density.
Main Results:
- A high prevalence of artifacts (97.1%) was observed across all images.
- Motion artifact was the most frequent (96.3%), followed by blinking (51.9%) and loss of focus (25.1%).
- MS patients showed a higher likelihood of artifacts, particularly motion artifacts, associated with longer disease duration and optic neuritis history. Artifacts were found to underestimate vessel density measurements.
Conclusions:
- Artifacts are prevalent in OCT-A imaging with the Spectralis® prototype and can significantly impact quantitative vascular density metrics.
- Careful image review for artifacts is essential before drawing conclusions from OCT-A studies, especially in MS populations.
- Further research may be needed to refine artifact detection and mitigation strategies in OCT-A.
More Related Videos
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Multiple Allele Traits
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Multiple Regression
Farmers can use multiple regression to determine the crop yield based on more than one factor, such as water availability, fertilizer, soil properties, etc. Here, the crop yield is the response or dependent variable as it depends on the other independent variables. The analysis requires the construction of a scatter plot...