Technical Note: Evaluation of kV CBCT enhancement using a liver-specific contrast agent for stereotactic body
John D Lincoln1, David Parsons2, Sharon E Clarke1,3,4
1Department of Physics and Atmospheric Science, Dalhousie University, Halifax, B3H 4R2, Canada.
Purpose:
To evaluate possible use for cone-beam computed tomography (CBCT) guidance, this phantom study evaluated the contrast enhancement provided by Gadoxetate Disodium (Primovist® CAN/EU, or Eovist® USA, Bayer Healthcare, Leverkusen, Germany), a contrast agent that is taken up selectively by liver cells and is retained for up to an hour. Image quality from CBCT was benchmarked against helical fan-beam computed tomography for two phantom geometries.
Methods And Materials:
Concentrations were diluted to 0.0125-0.1 mmol per kilogram of body weight (mmol/kg) corresponding to expected physiological concentrations in the liver. Kilovoltage CBCT imaging parameters of x-ray tube potential, current, and filtration were investigated using clinically available options on a TrueBeam STx linear accelerator CBCT platform. Two phantoms were created, a cylindrical idealized imaging geometry and an ellipsoidal more realistic abdominal geometry. All parameters were optimized according to the contrast-to-noise ratio (CNR) image quality metric, as a function of concentration, following the Rose criterion for CNR.
Results:
Acceptable CNR was defined as greater than or equal to three, in accordance with the Rose criterion for CNR. These were found in a range of expected liver concentrations of 0.025-0.1 mmol/kg for a tube potential of 100 kVp, half-fan bowtie filtration and tube currents giving exposures between 2025 and 5085 mAs. Linear correlations were found for all CNR as a function of concentration, in agreement with the literature.
Conclusion:
Based on this phantom study, with appropriate selection of imaging protocol, Gadoxetate Disodium may provide useful liver CBCT enhancement at physiologically achievable liver concentrations.
Related Concept Videos
Self-Evaluation: Self-Enhancement and Self-Verification
Biological Effects of Radiation
Cell Specific Gene Expression
Gene Therapy
Radiation: Applications
The average...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...