Related Experiment Video
Updated: Jan 30, 2026

Author Spotlight: Investigating HR-Dependent Cardiac Function in Mouse Models Through a Novel Atrial-Pacing Approach
Published on: July 21, 2023
Preliminary experience with the multisensor HeartLogic algorithm for heart failure monitoring: a retrospective case
Alessandro Capucci1, Luca Santini2, Stefano Favale3
1Clinica di Cardiologia e Aritmologia, Università Politecnica delle Marche, "Ospedali Riuniti", Via Conca, 71, 60020, Ancona, Italy.
Insights
The HeartLogic algorithm, using implantable cardioverter defibrillator (ICD) sensors, effectively predicts heart failure (HF) decompensation. This remote monitoring tool shows promise in managing HF patients and preventing hospitalizations.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Devices
Background:
- Heart failure (HF) management relies on timely detection of decompensation.
- Remote monitoring of HF patients is crucial for proactive care.
- Implantable cardioverter defibrillator (ICD) devices offer potential for integrated HF monitoring.
Purpose of the Study:
- To evaluate the clinical utility of the HeartLogic algorithm for remote HF monitoring.
- To assess the HeartLogic index's ability to predict HF decompensation events.
- To describe real-world post-implantation data of HeartLogic in HF patients.
Main Methods:
- Retrospective analysis of data from 58 patients with ICDs and cardiac resynchronization therapy (CRT-ICD).
- Activation of the HeartLogic feature on a remote monitoring platform.
- Analysis of sensor data (heart rate, sounds, impedance, respiration, activity) and the HeartLogic index against clinical events.
Main Results:
- HeartLogic alerts preceded 5 HF hospitalizations and 5 unplanned HF visits in 16 patients over 24 person-years.
- Alerts provided a median early warning time, longer for hospitalizations than minor deteriorations.
- Changes in heart sound amplitude were detected in all alert cases.
Conclusions:
- The HeartLogic algorithm demonstrates potential for detecting gradual HF worsening.
- This tool may aid in stratifying the risk of HF decompensation.
- Real-world clinical data supports HeartLogic's utility in HF patient management.
Aims:
In the Multisensor Chronic Evaluation in Ambulatory Heart Failure Patients study, a novel algorithm for heart failure (HF) monitoring was implemented. The HeartLogic (Boston Scientific) index combines data from multiple implantable cardioverter defibrillator (ICD)-based sensors and has proved to be a sensitive and timely predictor of impending HF decompensation. The remote monitoring of HF patients by means of HeartLogic has never been described in clinical practice. We report post-implantation data collected from sensors, the combined index, and their association with clinical events during follow-up in a group of patients who received a HeartLogic-enabled device in clinical practice.
Methods And Results:
Patients with ICD and cardiac resynchronization therapy ICD were remotely monitored. In December 2017, the HeartLogic feature was activated on the remote monitoring platform, and multiple ICD-based sensor data collected since device implantation were made available: HeartLogic index, heart rate, heart sounds, thoracic impedance, respiration, and activity. Their association with clinical events was retrospectively analysed. Data from 58 patients were analysed. During a mean follow-up of 5 ± 3 months, the HeartLogic index crossed the threshold value (set by default to 16) 24 times (over 24 person-years, 0.99 alerts/patient-year) in 16 patients. HeartLogic alerts preceded five HF hospitalizations and five unplanned in-office visits for HF. Symptoms or signs of HF were also reported at the time of five scheduled visits. The median early warning time and the time spent in alert were longer in the case of hospitalizations than in the case of minor events of clinical deterioration of HF. HeartLogic contributing sensors detected changes in heart sound amplitude (increased third sound and decreased first sound) in all cases of alerts. Patients with HeartLogic alerts during the observation period had higher New York Heart Association class (P = 0.025) and lower ejection fraction (P = 0.016) at the time of activation.
Conclusions:
Our retrospective analysis indicates that the HeartLogic algorithm might be useful to detect gradual worsening of HF and to stratify risk of HF decompensation.