Spectrophotometric determination of iodate in iodised salt by flow injection analysis
Ali Mohammad Haji Shabani1, Peter S Ellis2, Ian D McKelvie3
1Department of Chemistry, Yazd University, Yazd, Iran; Water Studies Centre, School of Chemistry, P.O. Box 23, Monash University, Victoria 3800, Australia.
Abstract:
A simple and sensitive flow injection (FI) method for the determination of iodate is proposed. The method is based on the reaction of iodate with hydroxylamine in acidic solution. Sulfanilamide is diazotised by the nascent nitrite and the diazonium ion produced is then coupled with N-(1-naphthyl)ethyenediamine in hydrochloric acid medium to form an azo dye which is measured spectrophotometrically. The calibration graph for iodate is linear in the range of 0.1-30mgL-1 with a correlation coefficient of 0.9992. The limit of detection and relative standard deviation are 0.02mgL-1 and 1.2% (5mgL-1, n=8). The method has been applied to the determination of iodate in table salts and accuracy was assessed through recovery experiments and independent analysis by a conventional titrimetric method.
More Related Videos
Related Concept Videos
Determining the pH of Salt Solutions
Responses to Salt Stress
Aryldiazonium Salts to Azo Dyes: Diazo Coupling
Determining Order of Reaction
Determination
Qualitative Analysis
For instance, group IV...