Accelerated multi-snapshot free-breathing mapping based on the dual refocusing echo acquisition mode technique
Teresa Rincón-Domínguez1,2, Anne Menini3, Ana Beatriz Solana3
1Technische Universität München, Munich, Germany.
Background:
Field inhomogeneities in MRI caused by interactions between the radiofrequency field and the patient anatomy can lead to artifacts and contrast variations, consequently degrading the overall image quality and thereby compromising diagnostic value of the images.
Purpose:
To develop an efficient free-breathing and motion-robust mapping method that allows for the investigation of spatial homogeneity of the transmitted radiofrequency field in the myocardium at 3.0T. Three joint approaches are used to adapt the dual refocusing echo acquisition mode (DREAM) sequence for cardiac applications: (1) electrocardiograph triggering; (2) a multi-snapshot undersampling scheme, which relies on the Golden Ratio, to accelerate the acquisition; and (3) motion-compensation based on low-resolution images acquired in each snapshot.
Study Type:
Prospective.
Phantom/Subjects:
Eurospin II T05 system, torso phantom, and five healthy volunteers.
Field Strength/Sequence:
3.0T/DREAM.
Assessment:
The proposed method was compared with the Bloch-Siegert shift (BSS) method and validated against the standard DREAM sequence. Cardiac maps were obtained in free-breathing and breath-hold as a proof of concept of the in vivo performance of the proposed method.
Statistical Tests:
Mean and standard deviation (SD) values were analyzed for six standard regions of interest within the myocardium. Repeatability was assessed in terms of SD and coefficient of variation.
Results:
Phantom results indicated low deviation from the BSS method (mean difference = 3%). Equivalent distributions for free-breathing and breath-hold in vivo experiments demonstrated the motion robustness of this method with good repeatability (SD < 0.05). The amount of variations was found to be 26% over the myocardium within a short axis slice.
Data Conclusion:
The feasibility of a cardiac mapping method with high spatial resolution in a reduced scan time per trigger was demonstrated. The free-breathing characteristic could be beneficial to determine shim components for multi-channel systems, currently limited to two for a single breath-hold.
Level Of Evidence:
1 Technical Efficacy: Stage 1 J. Magn. Reson. Imaging 2019;49:499-507.
More Related Videos
08:35Oxygenation-sensitive Cardiac MRI with Vasoactive Breathing Maneuvers for the Non-invasive Assessment of Coronary Microvascular Dysfunction
Published on: August 17, 2022
06:29Cardiac Magnetic Resonance for the Evaluation of Suspected Cardiac Thrombus: Conventional and Emerging Techniques
Published on: June 11, 2019
Related Concept Videos
Echo
Imagine the sound is reflected back to the ears. Assuming that the source is very close to the human, the difference between hearing the two sounds—the emitted sound and the reflected sound—may be more than the minimum time for perceiving distinct sounds. If this is the case,...
Measuring Acceleration Due to Gravity
A simple pendulum can be described as a point mass and a string. Meanwhile, a physical pendulum is any object whose oscillations are similar to a simple pendulum, but cannot be modeled as a point mass on a string because its mass is distributed over a larger area. The behavior of a physical pendulum can be modeled using the principles of...
Breathing
Dreaming
Lucid Dreaming
Studies have shown...
What is a Mode?
There can be more than one mode in a data set if multiple values have the same highest frequency. For instance, suppose that the Statistics exam scores of 20 students are: 50; 53; 59; 59; 63; 63; 72; 72; 72; 72; 72; 76; 78; 81; 83; 84; 84; 84; 90; 93. Here, the mode is 72, as it occurs most frequently, five times.
A data set with two modes is called bimodal. For example,...