Dosimetry Effects Caused by Unilateral and Bilateral Hip Prostheses: A Monte Carlo Case Study in Megavoltage Photon
Courage Mahuvava1, Frederik Carl Phillipus Du Plessis1
1Department of Medical Physics, Faculty of Health Sciences, University of the Free State, Bloemfontein, South Africa.
Background:
Hip prostheses (HPs) are routinely used in hip augmentation to replace painful or dysfunctional hip joints. However, high-density and high-atomic-number (Z) inserts may cause dose perturbations in the target volume and interface regions.
Aim:
To evaluate the dosimetric influence of various HPs during megavoltage conformal radiotherapy (RT) of the prostate using Monte Carlo (MC) simulations.
Materials And Methods:
BEAMnrc and DOSXYZnrc MC user-codes were respectively used to simulate the linac head and to calculate 3D absorbed dose distributions in a computed tomography (CT)-based phantom. A novel technique was used to synthetically introduce HPs into the raw patient CT dataset. The prosthesis materials evaluated were stainless steel (SS316L), titanium (Ti6Al4V), and ultra-high-molecular-weight polyethylene (UHMWPE). Four, five, and six conformal photon fields of 6-20 MV were used.
Results:
The absorbed dose within and beyond metallic prostheses dropped significantly due to beam attenuation. For bilateral HPs, the target dose reduction ranged up to 23% and 17% for SS316L and Ti6Al4V, respectively. For unilateral HP, the respective dose reductions were 19% and 12%. Dose enhancement was always <1% for UHMWPE. The 6-field plan produced the best target coverage. Up to 38% dose increase was found at the bone-SS316L proximal interface.
Conclusions:
The novel technique used enabled the complete exclusion of metal artifacts in the CT dataset. High-energy plans with more oblique beams can help minimize dose attenuation through HPs. Shadowing and interface effects are density dependent and greatest for SS316L, while UHMWPE poses negligible dose perturbation.
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Bonding in Metals
Metallic Solids
All metallic solids exhibit high thermal and electrical conductivity, metallic luster, and malleability....
Alkali Metals
Table 1: Properties of the alkali metals
Properties of Transition Metals