Imaging and Dosimetric Study on Direct Flat-Panel Detector-Based Digital Mammography System
Reena Sharma1,2, S D Sharma1,2, P S Sarkar2,3
1Radiological Physics and Advisory Division, Bhabha Atomic Research Centre, Mumbai, Maharashtra, India.
Introduction:
Image quality of digital mammography system is generally defined by three primary physical parameters, namely, contrast, resolution, and noise. Quantification of these metrics can be done by measuring objective image quality parameters defined as contrast-to-noise ratio (CNR), modulation transfer function (MTF), and noise power spectra (NPS).
Materials And Methods:
In the present study, various imaging metrics such as CNR, contrast detail resolution, MTF, and NPS were evaluated for a direct flat-panel detector-based digital mammography system following the European Guidelines. Furthermore, system performance relating to both image quality and doses were evaluated using figure of merit (FOM) in terms of CNR2/mean glandular dose (MGD) under automatic exposure control (AEC) and clinically used OPDOSE operating mode.
Results And Conclusion:
Under AEC mode, FOM values for the 4.5 cm thick BARC polymethyl methacrylate (PMMA) phantom were found to be 15.02, 15.88, and 19.82 at Mo/Mo, Mo/Rh, and W/Rh target/filter (T/F), respectively. Under OPDOSE mode, FOM values were found to 65.32, 11.80, and 1.14 for the BARC PMMA phantom thickness of 2, 4.5, and 8 cm, respectively. Under OPDOSE mode, the calculated MGD values for three Computerized Imaging Reference Systems slab phantoms having total thickness of 7 cm were observed to be 3.03, 2.32, and 1.75 mGy with glandular/adipose tissue compositions of 70/30, 50/50, and 30/70, respectively, whereas for the 2-8-cm thick BARC PMMA phantom, the calculated MGDs were found to be in the range of 0.57-3.32 mGy. All the calculated MGDs values were found to be lower than the acceptable level of dose limits provided in European Guidelines.
Related Concept Videos
Wood Panel Products
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Flat Belts: Problem Solving
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II
Frictional Forces on Flat Belts