All-polyethylene tibial components in young patients have stable fixation; a comparison RSA study
David Campbell1, Stuart Callary2, John Field2
1Wakefield Orthopaedic Clinic, 270 Wakefield Street, Adelaide, South Australia, Australia.
Background:
All-polyethylene (AP) tibial components in patients aged greater than 60 years have stable tibial migration patterns and favorable survival rates when compared to identical Metal-backed (MB) designs. Tibial component migration in younger patients has not been reported. The aim of this study was to examine the migration characteristics of patients aged less than 60 years compared to a previous cohort of AP and MB tibial components of identical design in older patients.
Methods:
A prospective consecutive study examined tibial component migration in 21 patients aged less than 60 years undergoing a cemented total knee arthroplasty with an AP tibial component by radiostereometric analysis (RSA) to 24 months. Results were compared to the authors' previous series of 21 patients aged greater than 60 years that were randomized to either an AP or MB tibial component.
Results:
Both age groups of patients implanted with an AP component had stable migration patterns with no patient having greater than 0.2° rotation or 0.2 mm maximum total point motion. Five of 11 MB tibial components displayed continued migration between one and two years. Subsidence was similar in all groups, whilst maximum total point motion was greater for the MB cohort (0.34 mm, 0.33 mm, 0.61 mm; AP <60, AP >60, MB).
Conclusions:
Young patients implanted with an AP tibial component had stable tibial migration patterns comparable to older patients with the same AP implant. Regardless of age, AP tibial components were at least as stable as MB tibial components.
Level Of Evidence:
Level II, Prospective comparative study.
More Related Videos
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Fixation and Sectioning
The simplest type of preparation is the wet mount, in which the specimen is placed in a drop of liquid on the slide. A liquid specimen can be directly deposited on the slide using a dropper. Solid specimens, such as skin scraping, can be placed on the slide before adding a drop of liquid to prepare the wet mount. Sometimes the liquid is simply water, but stains are often added...
Components of Stress
Interestingly, the hidden cube faces also experience these stresses, equal and...
Components of Language
Carbon-dioxide Fixation
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...