Assessing zinc from a nail clipping using mono-energetic portable X-ray fluorescence
David E B Fleming1, Samantha L Crook1, Colby T Evans1
1Physics Department, Mount Allison University, Sackville, NB, Canada.
Abstract:
A mono-energetic X-ray beam from a portable X-ray fluorescence device was used to excite characteristic X-rays from zinc in a series of nail clipping phantoms. Twenty nail clipping phantoms having equal zinc concentrations of ~40 µg/g, but with different physical characteristics, were measured individually for 300 s using a small diameter (~1 mm) X-ray beam. Energy spectra obtained from the measurements were analyzed using PyMca software. Zinc signal size varied widely between the different clippings, with a relative standard deviation of 41% observed in the combined signal from zinc Kα and Kβ characteristic X-rays. Three different normalization approaches were introduced to account for variation in the amounts of sample interrogated by the X-ray beam. All three approaches produced similar results, and successfully reduced the relative standard deviation to between 12% and 13%. A clear trend was still observed, however, between the normalized zinc signal and the thickness of clipping measured. To account for this effect, normalized signals were adjusted to calculate "thickness-corrected" values. The relative standard deviation of these thickness-corrected values was 6.2%. Reproducibility of measurement from individual clippings was excellent, with relative standard deviations on the order of 1%, with or without normalization. Overall, this new method of measuring zinc in nail shows promise for the assessment of zinc status in humans using a portable device. The method is sensitive, rapid, and requires only a single nail clipping.
Related Concept Videos
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...
Energetics of Solution Formation
When the strengths of the intermolecular forces of attraction between solute and solvent species in a solution are no different than those present in the separated components, the solution is formed with no accompanying energy change. Formation of the solution requires the solute–solute and solvent–solvent...
Accessory Structures of the Skin: Nails
The main components of a nail include the following.
Nail Plate: The nail plate is the visible portion of the nail that extends beyond the fingertips or toes. It is a hard, translucent...
X-ray Imaging
X-ray Diffraction of Biological Samples
According to Bragg's law, when X-rays strike the sample positioned on a stage, the rays are scattered by the electron clouds around the sample atoms. The X-ray diffraction or scattering is caused by constructive interference of the X-ray waves that reflect off the internal...
Radiological Investigation I: X-ray and CT