Systematic overestimation of human serum albumin by capillary zone electrophoresis method due to monoclonal
Maël Padelli1, Théo Labouret2, Marion Labarre2
1Department of Biochemistry and Pharmaco-Toxicology, University Hospital of Brest, Brest, France; Univ Brest, LIEN, F-29200 Brest, France.
Background:
The capillary zone electrophoresis method of albumin measurement is frequently used in monoclonal gammopathy patients but some studies suggest poor performances of the method in this population. The aim of this study was to analyse the impact of serum monoclonal immunoglobulins on human serum albumin determination by capillary zone electrophoresis method compared to other available methods.
Method:
We prospectively measured albumin in 100 freshly collected non-frozen serum samples in a monoclonal gammopathy patients population, by using four different methods: the capillary zone electrophoresis method, the bromocresol purple dye method, the nephelometric method and the turbidimetric method. Differences in albumin values between the different methods were analysed with respect to serum monoclonal immunoglobulin concentration. These differences were further investigated by measuring albumin levels in human serum samples spiked with exogenous monoclonal immunoglobulins.
Results:
Human serum albumin difference values between capillary zone electrophoresis compared to immunonephelometry method are significantly correlated with increasing monoclonal immunoglobulins concentrations: regression analyses revealed a correlation coefficient r2 = 0.60 and a slope of 0.14 (0.12-0.17, 95% confidence interval). The capillary zone electrophoresis method overestimated serum albumin levels by up to 67% (12 g/L) when monoclonal immunoglobulin level was 63 g/L. The determination of albumin levels in human serum samples spiked with exogenous monoclonal immunoglobulins showed an overestimation of human serum albumin measurement by the capillary zone electrophoresis method proportional to the amount of monoclonal immunoglobulin added in the serum with a slope of 0.19 (0.18-0.20, 95% confidence interval).
Conclusion:
Monoclonal immunoglobulins directly interfere with serum albumin measurement by the capillary zone electrophoresis method leading to a systematic overestimation of serum albumin concentrations proportional to the serum monoclonal immunoglobulin level.
More Related Videos
10:05Large-scale Top-down Proteomics Using Capillary Zone Electrophoresis Tandem Mass Spectrometry
Published on: October 24, 2018
07:54Analysis of AtHIRD11 Intrinsic Disorder and Binding Towards Metal Ions by Capillary Gel Electrophoresis and Affinity Capillary Electrophoresis
Published on: August 22, 2018
Related Concept Videos
Capillary Electrophoresis: Instrumentation
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Systematic Sampling Method
Systematic sampling is one of the simplest methods...
Hybrid Zones
Interference and Diffraction
RNA Interference
This process occurs naturally in cells, often through the activity of genomically-encoded microRNAs. Researchers can take advantage of this mechanism by introducing synthetic RNAs to deactivate specific genes for research or therapeutic purposes. For example, RNAi could be used...