Ex vivo performance comparison of three percutaneous biopsy needle systems
Miltiadis Krokidis1,2,3, Martin Hungerbühler4,5, Ekkehard Hewer6
1Department of Radiology, Cambridge University Hospitals NHS Foundation Trust, Cambridge Biomedical Campus, Cambridge, UK. mkrokidis@hotmail.com.
Purpose:
The aim of this study is to identify the micro-mechanical characteristics that define biopsy performance in normal ex vivo animal organs.
Materials And Methods:
Three biopsy systems with differences of needle external diameter, tray height and effective tray length were assessed. Sampling was performed in porcine liver and kidneys with commercially labelled 14G, 16G and 18G, using 2-cm throw needle systems. Five samples were obtained per needle size and per organ, and the experiment was repeated twice for a total of 90 biopsy cores. Samples were analysed and compared in terms of macroscopic aspect, sample length, weight and tissue architecture.
Results:
The system with the longest effective needle tray (system A) has shown significant superiority (p < 0.001) versus the other systems regarding the mean weight of tissue obtained for all needle sizes. Furthermore, the 14G needle of system A has shown superiority regarding the number of portal spaces and the 16G regarding the length of kidney fragments obtained.
Conclusion:
The outcomes obtained with the different biopsy systems were not equal. The micro-mechanical characteristic that appears to influence the quantity and quality of the obtained tissue is the effective needle tray length and not the needle external diameter or the needle tray height. This information should be taken into account in the future design of biopsy needle systems, particularly when potentially used in the assessment of biomarkers and the characterisation of tumour micro-environment.
Key Points:
• The amount of obtained tissue mass is not the same among percutaneous biopsy needle systems. • There are different micro-mechanical characteristics that condition the amount of obtained tissue. • The micro-mechanical characteristic that offers more tissue mass for the same calibre is the effective length of the needle tray.
More Related Videos
11:27Preparation and Respirometric Assessment of Mitochondria Isolated from Skeletal Muscle Tissue Obtained by Percutaneous Needle Biopsy
Published on: February 7, 2015
10:19Substernal Thyroid Biopsy Using Endobronchial Ultrasound-guided Transbronchial Needle Aspiration
Published on: November 10, 2014
Related Concept Videos
Drug Product Performance: In Vitro–In Vivo Correlation
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Second Order systems II
Evolutionary Relationships through Genome Comparisons
First Order Systems
When a first-order system is subjected to a unit-step input, its response is characterized by its transfer function. By applying the Laplace transform of the unit-step input to the transfer function, expanding the...