Related Experiment Video
Updated: Jan 30, 2026

In Silico Clinical Trials for Cardiovascular Disease
Published on: May 27, 2022
Evaluation of Digital Breast Tomosynthesis as Replacement of Full-Field Digital Mammography Using an In Silico
Aldo Badano1, Christian G Graff1, Andreu Badal1
1Division of Imaging, Diagnostics, and Software Reliability, Office of Science and Engineering Laboratories, Center for Devices and Radiological Health, US Food and Drug Administration, Silver Spring, Maryland.
Importance:
Expensive and lengthy clinical trials can delay regulatory evaluation of innovative technologies, affecting patient access to high-quality medical products. Simulation is increasingly being used in product development but rarely in regulatory applications.
Objectives:
To conduct a computer-simulated imaging trial evaluating digital breast tomosynthesis (DBT) as a replacement for digital mammography (DM) and to compare the results with a comparative clinical trial.
Design, Setting, And Participants:
The simulated Virtual Imaging Clinical Trial for Regulatory Evaluation (VICTRE) trial was designed to replicate a clinical trial that used human patients and radiologists. Images obtained with in silico versions of DM and DBT systems via fast Monte Carlo x-ray transport were interpreted by a computational reader detecting the presence of lesions. A total of 2986 synthetic image-based virtual patients with breast sizes and radiographic densities representative of a screening population and compressed thicknesses from 3.5 to 6 cm were generated using an analytic approach in which anatomical structures are randomly created within a predefined breast volume and compressed in the craniocaudal orientation. A positive cohort contained a digitally inserted microcalcification cluster or spiculated mass.
Main Outcomes And Measures:
The trial end point was the difference in area under the receiver operating characteristic curve between modalities for lesion detection. The trial was sized for an SE of 0.01 in the change in area under the curve (AUC), half the uncertainty in the comparative clinical trial.
Results:
In this trial, computational readers analyzed 31 055 DM and 27 960 DBT cases from 2986 virtual patients with the following Breast Imaging Reporting and Data System densities: 286 (9.6%) extremely dense, 1200 (40.2%) heterogeneously dense, 1200 (40.2%) scattered fibroglandular densities, and 300 (10.0%) almost entirely fat. The mean (SE) change in AUC was 0.0587 (0.0062) (P < .001) in favor of DBT. The change in AUC was larger for masses (mean [SE], 0.0903 [0.008]) than for calcifications (mean [SE], 0.0268 [0.004]), which was consistent with the findings of the comparative trial (mean [SE], 0.065 [0.017] for masses and -0.047 [0.032] for calcifications).
Conclusions And Relevance:
The results of the simulated VICTRE trial are consistent with the performance seen in the comparative trial. While further research is needed to assess the generalizability of these findings, in silico imaging trials represent a viable source of regulatory evidence for imaging devices.
More Related Videos
Related Concept Videos
Social Foundations of Self IV: Self in Digital Communication
Clinical Trials: Overview
Trial and Error and Algorithm
Clinical Trials
There are four phases in a clinical trial. A phase one...
Continuous Renal Replacement Therapy
Crystal Field Theory - Octahedral Complexes
To explain the observed behavior of transition metal complexes (such as colors), a model involving electrostatic interactions between the electrons from the ligands and the electrons in the unhybridized d orbitals of the central metal atom has been developed. This electrostatic model is crystal field theory (CFT). It helps to understand, interpret, and predict the colors, magnetic behavior, and some structures of coordination compounds of transition metals.
CFT focuses on...