Long-Term Outcomes of Vibroplasty Coupler Implantations to Treat Mixed/Conductive Hearing Loss
Thomas Zahnert1, Robert Mlynski2,3, Hubert Löwenheim4
1Department of Medicine, Clinic of Otorhinolaryngology, Head and Neck Surgery, Technical University of Dresden, Dresden, Germany, Thomas.Zahnert@uniklinikum-dresden.de.
Objective:
To evaluate the long-term safety and performance of four different vibroplasty couplers (round window, oval window, CliP and Bell coupler) in combination with an active middle ear implant.
Methods:
This was a multicentre, prospective, long-term study including 5 German hospitals. Thirty adult subjects suffering from conductive or mixed hearing loss were initially enrolled for the study, 24 of these were included in the final analysis with up to 36 months of postsurgical follow-up data. Bone conduction and air conduction were measured pre- and postoperatively to evalu ate safety. Postoperative aided sound field thresholds and Freiburger monosyllable word recognition scores were compared to unaided pre-implantation results to confirm performance. Additional speech tests compared postoperative unaided with aided results. To determine patient satisfaction, an established quality-of-life questionnaire developed for conventional hearing aid usage was administered to all subjects.
Results:
Mean postoperative bone conduction thresholds remained stable throughout the whole study period. Mean functional gain for all couplers investigated was 38.5 ± 11.4 dB HL (12 months) and 38.8 ± 12.5 dB HL (36 months). Mean word recognition scores at 65 dB SPL increased from 2.9% in the unaided by 64.2% to 67.1% in the aided situation. The mean postoperative speech reception in quiet (or 50% understanding of words in sentences) shows a speech intelligibility improvement at 36 months of 17.8 ± 12.4 dB SPL over the unaided condition. The signal-to-noise ratio (SNR) improved by 5.9 ± 7.2 dB SNR over the unaided condition. High subjective device satisfaction was reflected by the International Inventory for Hearing Aids scored very positively.
Conclusion:
A significant improvement was seen with all couplers, and audiological performance did not significantly differ between 12 and 36 months after surgery.
More Related Videos
Related Concept Videos
Hearing
Line Loss
Line loss impacts power delivery efficiency in a balanced three-phase circuit. The symmetry in such a circuit simplifies the...
Predicting Reaction Outcomes
Reducing Line Loss
With a step-up transformer at the source, the voltage is increased, thereby reducing the current in the transmission lines since power loss in...
Outcomes of Glycolysis
Cellular respiration can occur aerobically (with oxygen) or anaerobically (without oxygen). In the presence of oxygen, cellular respiration starts with glycolysis and continues with pyruvate...
Long-term Depression