Performance characteristics of the VIDAS® ANTI-HEV IgM and IgG assays
Florence Abravanel1, Nadège Goutagny2, Raphaelle Joffray3
1CHU Toulouse, Hôpital Purpan, Laboratoire de virologie, National Reference Center for Hepatitis E, F-31300, France; INSERM, U1043, Centre de Physiopathologie de Toulouse Purpan, Toulouse, F-31300, France.
Background:
Several unautomated anti-HEV diagnostic tests are presently available.
Objective:
We have evaluated the performance of the new automated VIDAS® ANTI-HEV IgM and IgG assays.
Study Design:
We assessed the reproducibility and cross-reactivity of both VIDAS assays and the analytical sensitivity and linearity of the VIDAS IgG assay. We also tested the VIDAS and comparator assays Wantai IgG and IgM on immunocompetent and immunocompromised patients. Data were analysed according to the infectious profile, with samples from viremic phase (HEV RNA/IgM positive) and post-viremic phase (HEV RNA negative, IgM positive) infections, and uninfected patients (HEV RNA/IgM negative).
Results:
Within-run reproducibility was <10% and between-run reproducibility was <12% for both assays. We found no cross-reactivity, except for the VIDAS IgG assay in some patients with HBV (1/10) or malaria (3/23) infections and for the VIDAS IgM assay in some HIV-infected patients (1/10). The VIDAS IgG assay was linear over 0.10-10.0 U/mL. Analytical sensitivity of the IgG assay was 0.71 IU/ml (probit analysis). The clinical sensitivity of the VIDAS IgM assay was 97.65% for viremic samples (83/85) and 59.15% (42/71) for post-viremic samples from immunocompetent patients. It was 78.95% (45/57) for acute phase samples and 77.78% (28/36) for post-viremic samples from immunocompromised patients. Specificity was excellent (>99%) in both populations.
Conclusion:
The analytical and clinical performance of the new VIDAS® ANTI-HEV assays was excellent. These rapid, automated assays for detecting HEV antibodies will strengthen the arsenal for diagnosing HEV infections.
Related Concept Videos
Transcytosis of IgG
IgG molecules from a mother undergo transcytosis starting around 13 weeks of gestation. The amount of IgG transferred and entering the fetal blood circulation increases with...
Characteristics of Life
Characteristics of Fluids
Characteristics of Fluids
Fluids, which include both liquids and gases, are substances that deform continuously under shearing stress. For example, water and oil are liquids with molecules that can...
Characteristics of BJT
For input characteristics, the base-emitter voltage is varied, maintaining a constant collector-emitter voltage. This setup reveals a Shockley-type dependence of the collector current on...
Characteristics of JFET
The core of a JFET's operation is controlling drain current by modulating the gate-source voltage. When the drain and gate voltage are set to zero, the JFET exhibits no net current flow, representing a state of equilibrium. The drain current increases linearly as the...