Single- and dual-energy CT pulmonary angiography using second- and third-generation dual-source CT systems:
Lukas Lenga1, Franziska Trapp2, Moritz H Albrecht3
1Division of Experimental Imaging, Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Theodor-Stern-Kai 7, 60590, Frankfurt, Germany. lukas.lenga@gmail.com.
Objectives:
To evaluate radiation exposure and image quality in matched patient cohorts for CT pulmonary angiography (CTPA) acquired in single- and dual-energy mode using second- and third-generation dual-source CT (DSCT) systems.
Methods:
We retrospectively included 200 patients (mean age, 65.5 years ± 15.7 years) with suspected pulmonary embolism-equally divided into four study groups (n = 50) and matched by gender and body mass index. CTPA was performed with vendor-predefined second-generation (group A, 100-kV single-energy computed tomography (SECT); group B, 80/Sn140-kV dual-energy computed tomography (DECT)) or third-generation DSCT (group C, 100-kV SECT; group D, 90/Sn150-kV DECT) devices. Radiation metrics were assessed using a normalized scan range of 27.5 cm. For objective image quality evaluation, dose-independent figure-of-merit (FOM) contrast-to-noise ratios (CNRs) were calculated. Subjective image analysis included ratings for overall image quality, reader confidence, and image artifacts using five-point Likert scales.
Results:
Calculations of the effective dose (ED) of radiation for a normalized scan range of 27.5 cm showed nonsignificant differences between SECT and DECT acquisitions for each scanner generation (p ≥ 0.253). The mean effective radiation dose was lower for third-generation groups C (1.5 mSv ± 0.8 mSv) and D (1.4 mSv ± 0.7 mSv) compared to second-generation groups A (2.5 mSv ± 0.9 mSv) and B (2.3 mSv ± 0.6 mSv) (both p ≤ 0.013). FOM-CNR measurements were highest for group D. Qualitative image parameters of overall image quality, reader confidence, and image artifacts showed nonsignificant differences among the four groups (p ≥ 0.162).
Conclusions:
Third-generation DSCT systems show lower radiation dose parameters for CTPA compared to second-generation DSCT. DECT can be performed with both scanner generations without radiation dose penalty or detrimental effects on image quality compared to SECT.
Key Points:
• Radiation exposure showed nonsignificant differences between SECT and DECT for both DSCT scanner devices. • Dual-energy CTPA provides equivalent image quality compared to standard image acquisition. • Subjective image quality assessment was similar among the four study groups.
More Related Videos
08:04Monitoring Tumor Metastases and Osteolytic Lesions with Bioluminescence and Micro CT Imaging
Published on: April 14, 2011
02:09Multi-modal Pulmonary Imaging: Using Complementary Information from CT and Hyperpolarized 129Xe MRI to Evaluate Lung Structure-Function
Published on: April 12, 2024
Related Concept Videos
Dual Nature of Electromagnetic (EM) Radiation
Wavelength is the distance between two consecutive peaks (the highest point) or troughs (the lowest point) in the wave. Frequency is the number of...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Radiological Investigation I: X-ray and CT
Biological Effects of Radiation