Pulse Wave Velocity Testing in the Baltimore Longitudinal Study of Aging
Published on: February 7, 2014
Pulse wave analysis using the Mobil-O-Graph, Arteriograph and Complior device: a comparative study
Dimitrios Benas1, Michalis Kornelakis1, Helen Triantafyllidi1
1a Faculty of Medicine, Attikon Hospital, 2nd Department of Cardiology, 2nd Department of Pathology , National and Kapodistrian University of Athens , Athens , Greece.
Insights
Three devices for measuring pulse wave velocity (PWV) and central systolic blood pressure (cSBP) showed acceptable agreement. However, the Mobil-O-Graph may underestimate these arterial stiffness markers.
Area of Science:
- Cardiovascular diagnostics
- Biomedical engineering
- Clinical physiology
Background:
- Pulse wave velocity (PWV) is a key indicator of arterial stiffness and holds significant prognostic value.
- Arterial stiffness assessment is crucial for cardiovascular risk stratification.
- Accurate measurement of PWV and central systolic blood pressure (cSBP) is essential for clinical decision-making.
Purpose of the Study:
- To compare the Arteriograph and Complior devices against the Mobil-O-Graph for the assessment of PWV and cSBP.
- To evaluate the agreement and potential discrepancies between these three commonly used devices.
Main Methods:
- A study involving 316 subjects was conducted to measure PWV and cSBP using Arteriograph, Complior, and Mobil-O-Graph.
- Statistical analyses included Bland-Altman analysis, intraclass correlation coefficient (ICC), and coefficient of variation (CV) to compare device readings.
- Data were analyzed to determine mean differences, limits of agreement, and reliability between the devices.
Main Results:
- Bland-Altman analysis revealed acceptable mean differences for PWV (0.5-0.6 m/s) and cSBP (3.8-9.2 mmHg) between the devices.
- High intraclass correlation coefficients (0.86-0.92) indicated good reliability for both PWV and cSBP measurements across devices.
- The Mobil-O-Graph tended to show slightly lower values for PWV and cSBP compared to the other devices.
Conclusions:
- The study demonstrates acceptable agreement among Arteriograph, Complior, and Mobil-O-Graph for pulse wave analysis markers.
- The Mobil-O-Graph may underestimate PWV and cSBP values, warranting consideration in clinical practice.
- Further research is recommended to validate these findings across diverse clinical settings and patient populations.
Purpose:
Pulse wave velocity (PWV) is a marker of arterial stiffness with major prognostic value. We compared Arteriograph and Complior devices with the Mobil-O-Graph for assessment of PWV and central systolic blood pressure (cSBP).
Materials And Methods:
We studied 316 consecutive subjects (age: 55 ± 14 years). For each individual, we measured PWV and cSBP with Arteriograph, Complior and Mobil-O-Graph and compared the readings. Differences in values among three devices were calculated for each measurement. We used Bland-Altman analysis, intraclass correlation coefficient (ICC) and coefficient of variation (CV).
Results:
Bland-Altman analysis indicated a mean difference for PWV: i.0.5 m/s (limits of agreement -1.4-2.4) between Complior and Mobil-O-Graph, ii.0.6 m/s (limits of agreement -1.4-2.6) between Arteriograph and Mobil-O-Graph. cSBP mean difference was 3.8 mmHg between Complior and Mobil-O-Graph (limits of agreement -12.5-20.1) and 9.2 mmHg between Arteriograph and Mobil-O-Graph (limits of agreement -7.6-26). ICC for PWV was 0.86 between Arteriograph and Mobil-O-Graph, 0.87 between Complior and Mobil-O-Graph and for cSBP 0.92 and 0.91 respectively. CV for PWV was 9.5% between Arteriograph and Mobil-O-Graph, 8.8% between Complior and Mobil-O-Graph. Respective values for cSBP were 6.8% and 5.1%.
Conclusion:
Our study shows acceptable agreement among the three devices regarding pulse wave analysis markers though Mobil-O-Graph appears to underestimate the values of these markers. Further studies are needed to explore the agreement between the 3 devices in various clinical settings and patient populations.