Somatosensory evoked potential: Preventing brachial plexus injury in transaxillary robotic surgery
Shuo Huang1, Meghan E Garstka1, Mohammed A Murcy1
1Department of Surgery, New Orleans, Louisiana.
Objectives/Hypothesis:
The potential for brachial plexopathy due to arm positioning is a major concern regarding the robotic transaxillary approach. Intraoperative nerve monitoring via somatosensory evoked potential (SSEP) has been suggested to prevent such injury. In this study, we examined the use of SSEP in detecting imminent brachial plexus traction during robotic transaxillary thyroid and parathyroid surgery.
Study Design:
Retrospective case series.
Methods:
A analysis was performed for all patients undergoing robotic transaxillary surgery with continuous intraoperative SSEP monitoring at a North American institution between 2015 and 2017. A significant intraoperative SSEP change was defined as a decrease in signal amplitude of >50% or an increase in latency of >10% from baseline established during preoperative positioning.
Results:
One hundred thirty-seven robotic transaxillary surgeries using SSEP monitoring were performed on 123 patients. Seven patients (5.1%) developed significant changes, with an average SSEP amplitude reduction of 73% ± 12% recorded at the signals' nadir. Immediate arm repositioning resulted in recovery of signals and complete return to baseline parameters in 14.3 ± 9.2 minutes. There was no difference in age (40.4 ± 9.4 years vs. 44.5 ± 13.4 years; P = .31) or body mass index (27.3 ± 3.7 kg/m2 vs. 26.9 ± 6.1 kg/m2 ; P = .79) between cases with and without SSEP change. Operative time was shorter for patients with significant SSEP change (131.6 ± 14.7 minutes vs. 146.5 ± 46.7 minutes; P = .048). There were no postoperative positional brachial plexus injuries.
Conclusions:
SSEP is a novel, safe, and reliable tool in detection of position-related brachial plexus neuropathy. Intraoperative monitoring using SSEP can play a vital role in early recognition and prevention of injury during robotic transaxillary thyroid and parathyroid surgery.
Level Of Evidence:
4 Laryngoscope, 129:2663-2668, 2019.
More Related Videos
09:14Surface Electromyographic Biofeedback as a Rehabilitation Tool for Patients with Global Brachial Plexus Injury Receiving Bionic Reconstruction
Published on: September 28, 2019
07:27Robotics in Surgery: A Modular Robotic Platform Driven Gastric Wedge Resection
Published on: February 7, 2025
Related Concept Videos
Acute Kidney Injury IV: Diagnostic Studies and Prevention
Spinal Nerves: Plexus I
The Cervical Plexus
The cervical plexus, formed by the anterior rami of the first four...
Spinal Nerves: Plexus II
The Lumbar Plexus
The lumbar plexus is situated within the lumbar region of the back and is primarily formed by the first four lumbar spinal nerves (L1 to L4). This plexus extends its branches into several nerves, including the...
Cancer Prevention
Some...
Sensory Perception: Organization of the Somatosensory System
The receptor level:
The receptor level is the first stage of sensation. It involves the detection of a stimulus by specialized sensory receptors. The stimulus must arrive within the receptor's receptive field. Next, the receptor converts the energy of the...
Somatosensory, Motor, and Association Cortex