Improving the prediction of the trabecular bone microarchitectural parameters using dental cone-beam computed
Rong-Ting He1, Ming-Gene Tu1,2, Heng-Li Huang1,3
1School of Dentistry, College of Dentistry, China Medical University, Taichung, 404, Taiwan, Republic of China.
Background:
In this study, we explored how various preprocessing approaches can be employed to enhance the capability of dental CBCT to accurately estimate trabecular bone microarchitectural parameters.
Methods:
In total, 30 bovine vertebrae cancellous bone specimens were used for in study. Voxel resolution 18-μm micro-computed tomography (micro-CT) and 100-μm dental CBCT were used to scan each specimen. Micro-CT images were used to calculate trabecular bone microarchitectural parameters; the results were set as the gold standard. Subsequently, before the dental CBCT images were converted into binary images to calculate trabecular bone microarchitectural parameters, three preprocessing approaches were used to process the dental CBCT images. For Group 1, no preprocessing approach was applied. For Group 2, images were sharpened and despeckable noises were removed. For Group 3, the function of local thresholding was added to Group 2 to form Group 3. For Group 4, the air pixels was removed from Group 3 to form Group 4. Subsequently, all images were imported into a software package to estimate trabecular bone microarchitectural parameters (bone volume fraction (BV/TV), trabecular thickness (TbTh), trabecular number (TbN), and trabecular separation (TbSp)). Finally, a paired t-test and a Pearson correlation test were performed to compare the capability of micro-CT with the capability of dental CBCT for estimating trabecular bone microarchitectural parameters.
Results:
Regardless of whether dental CBCT images underwent image preprocessing (Groups 1 to 4), the four trabecular bone microarchitectural parameters measured using dental CBCT images were significantly different from those measured using micro-CT images. However, after three image preprocessing approaches were applied to the dental CBCT images (Group 4), the BV/TV obtained using dental CBCT was highly positively correlated with that obtained using micro-CT (r = 0.87, p < 0.001); the correlation coefficient was greater than that of Group 1 (r = -0.15, p = 0.412), Group 2 (r = 0.16, p = 0.386), and Group 3 (r = 0.47, p = 0.006). After dental CBCT images underwent image preprocessing, the efficacy of using dental CBCT for estimating TbN and TbSp was enhanced.
Conclusions:
Image preprocessing approaches can be used to enhance the efficacy of using dental CBCT for predicting trabecular bone microarchitectural parameters.
More Related Videos
05:49Author Spotlight: Advancing CBCT and Digital Dental Image Integration with AI-Assisted Digitization
Published on: February 23, 2024
10:23Author Spotlight: Three-Dimensional Cephalometric Landmark Annotation Demonstration on Human Cone Beam Computed Tomography Scans
Published on: September 8, 2023
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Predicting Molecular Geometry
Deflection of a Beam
Singularity functions, described in an earlier lesson, are powerful mathematical tools that represent discontinuities within a function commonly encountered in structural loading...
Wave Parameters