Test-Retest Reliability of Isokinetic Wrist Strength and Proprioception Measurements
Context:
The evaluation of the wrist strength and proprioception gives clinicians and researchers information about effectiveness of their rehabilitation protocol or helps diagnosis of various neuromuscular and somatosensorial disorders. Isokinetic dynamometers are considered the gold standard for these evaluations. However, the studies about test-retest reliability of isokinetic dynamometer are inadequate.
Objective:
The purpose of this study was to determine the test-retest reliability of isokinetic wrist strength and proprioception measurements using the Cybex isokinetic dynamometer.
Design:
Test-retest reliability study.
Setting:
University laboratory.
Participants:
Thirty participants were enrolled (age 23.2 [2.8] y, height 171.1 [7] cm, weight 66.6 [11.6] kg) in this study.
Intervention:
Cybex isokinetic dynamometer was used for strength and proprioception measurements.
Main Outcome Measures:
Concentric flexion-extension strength test was performed at 90°/s angular velocity, and eccentric flexion-extension strength test was performed at 60°/s angular velocity. The proprioception of the wrist was assessed via active joint position sense. The 30° extension of the wrist, which is accepted as the functional position of the wrist, was selected as the targeted angle. The intraclass correlation coefficient (ICC2,1) method was used for test-retest analysis (P < .05).
Results:
The active joint position sense measurements of dominant (ICC2,1: .821) and nondominant (ICC2,1: .763) sides were found to have good test-retest reliability. Furthermore, with the exception of dominant eccentric extension strength (moderate reliability) (ICC2,1: .733), eccentric and concentric flexion (dominant: ICC2,1 = .890-.844; nondominant: ICC2,1 = .800-.898, respectively), and extension (dominant: ICC2,1 = .791 [concentric], nondominant: ICC2,1 = .791-.818, respectively) strength measurements of both sides were found to have good reliability.
Conclusions:
This study shows that the Cybex isokinetic dynamometer is a reliable method for measuring wrist strength and proprioception. Isokinetic dynamometers can be used clinically for diagnosis or rehabilitation in studies which contain wrist proprioception or strength measurements.
Related Concept Videos
Reliability and Validity
Acid Strength and Molecular Structure
In the absence of any leveling effect, the acid strength of binary compounds of hydrogen with nonmetals (A) increases as the H-A bond strength decreases down a group in the periodic table. For group 17, the order of increasing acidity is HF < HCl < HBr < HI. Likewise, for group 16, the order of increasing acid strength is H2O < H2S < H2Se < H2Te. Across a row in the periodic table, the acid strength of binary hydrogen compounds increases with increasing...
The Equilibrium Binding Constant and Binding Strength
Strength of Cement
For compressive strength tests, ASTM C 109-05 standards prescribe a cement-sand mix ratio of 1:2.75 and a water/cement ratio of 0.485 for making 2-inch cubes. These cubes are mixed, cast, and cured in saturated lime water at 23°C until testing. Flexural strength testing, outlined in...
Relation Between Tensile Strength and Compressive Strength of Concrete
Distribution Reliability and Automation