Radiation Planning Assistant - A Streamlined, Fully Automated Radiotherapy Treatment Planning System
Published on: April 11, 2018
Assessing the feasibility of adaptive planning for prostate radiotherapy using Smartadapt deformable image
Virginia Marin Anaya1, Jamie Fairfoul2
1Medical Physics Department, Biomedical Campus, Cambridge University Hospitals NHS Foundation Trust, Addenbrooke's Hospital, Hills Road, Cambridge CB2 0QQ, United Kingdom.
Background:
The purpose of this study is to assess the feasibility of adaptive planning for prostate cancer using SmartAdapt deformable image registration (DIR), in an effort to reduce the effect of patient-specific variations.
Methods:
18 prostate patients (74 Gy in 37 fractions) were selected. Each patient had a planning CT and cone-beam CT (CBCT) images acquired at different fractions. For each fraction, the planning CT was registered to the CBCT image dataset using SmartAdapt. A rigid registration was performed prior to the DIR. The manual contours on the CBCT images by the clinician were used as reference. For the geometric analysis, various evaluation metrics were used to compare the volumes for both rigid and DIR. For the dosimetric analysis, the initial treatment plan was recalculated on the smartadapted structures, using Eclipse. Several Dose-Volume Histogram (DVH) points were used for the evaluation.
Results:
The results of the geometric analysis indicate that DIR can result in worse evaluation metrics than rigid registration alone. In general, SmartAdapt tends to under-contour the volumes. The difference in DVH points between the deformed and the reference prostate was worse than the difference between the planning CT and the reference, overestimating the need for re-planning. However, there was a tendency for the deformed rectum to underestimate the need for re-planning.
Conclusions:
After cautious analysis of the results, it has been decided not to implement SmartAdapt clinically to trigger adaptive decisions.
Related Concept Videos
Plastic Deformations
Plastic Deformations
Sampling Plans
Random sampling is a method where each member of the population has an equal chance of being selected for the sample. It involves selecting individuals randomly, often using random number generators or lottery-type methods. For example, when analyzing the properties of a...
Temperature Dependent Deformation
Deformations in a Symmetric Member in Bending
When the member is segmented into tiny cubic elements, it is observed that the primary stress...
Deformation of Member under Multiple Loadings
In the case of a member with a variable cross-section, the strain is not constant but depends on the position. The deformation of an...