Prospective Multicenter Study of the Low-Profile Relay Stent-Graft in Patients with Thoracic Aortic Disease: The
Vicenç Riambau1, Rocco Giudice2, Piero Trabattoni3
1Hospital Clínic i Provincial de Barcelona, Barcelona, Spain.
Background:
To evaluate the early safety and clinical performance of the new low-profile RelayPro Thoracic Stent-Graft System in patients with thoracic aortic disease.
Methods:
This was an international, prospective, single-arm study in patients diagnosed with thoracic aorta disease (aneurysm, pseudoaneurysm, dissection, penetrating atherosclerotic ulcer, or intramural hematoma) and treated with a RelayPro stent-graft (in bare stent and/or nonbare stent configurations). The primary endpoints were freedom from aneurysm or dissection-related mortality and stent-graft performance.
Results:
A total of 31 patients were treated with the RelayPro thoracic stent-graft between 2014 and 2015 at 8 sites in Italy and Spain. Mean age was 72.1 (±10.2) years and 77% were male, 74% with hypertension, and 42% with a history of smoking. Twenty-four (77%) had aneurysms (fusiform in 46%, saccular in 42%, pseudoaneurysm in 12%); 5 (16%) had penetrating atherosclerotic ulcer; and 2 (6%) had chronic Type B dissection. Mean vascular access diameter was 9.1 mm (6-13 mm); 7 patients (23%) had vascular access of 7 mm or less. Technical success was 100% (primary, 90%; assisted primary, 10%). Freedom from aneurysm/dissection-related mortality through 30 days was 100%. Freedom from device-related major adverse events through 30 days was 94%. At 1 year, there was 1 (3%) type Ib and 1 (3%) type II endoleak, 1 (3%) nonaneurysm-related late death, and 1 (3%) secondary intervention (to correct type Ib endoleak).
Conclusions:
The RelayPro has a 3-4 French profile reduction to allow endovascular repair of thoracic aortic disease in patients with smaller anatomies. This study shows good initial stent-graft performance and a favorable early safety profile.
Related Concept Videos
Directional Relays
Bioavailability Study Design: Healthy Subjects Versus Patients
Blood Studies for Cardiovascular System III: Serum Lipid Profile
Serum lipids are fats and fatty substances in the blood and are crucial for various bodily functions, including energy storage, cellular structure, and hormone production. Serum lipids consist of cholesterol, triglycerides, and phospholipids.
Cholesterol is a soft, fat-like substance found in all body cells. It is crucial for producing hormones, vitamin D, and substances that aid...
Case Studies
Overcurrent Relays
Instantaneous overcurrent relays activate immediately when the input current exceeds a predetermined value, known as the pickup current, instantly energizing the circuit breaker trip coil. This rapid response is vital for addressing severe faults quickly.
Time-delay overcurrent relays, on the other...
Differential Relays