Preliminary Evaluation of a Mobile Device for Dark Adaptation Measurement
1Schepens Eye Research Institute of Mass Eye & Ear, Harvard Medical School, Boston, MA, USA.
Purpose:
We evaluated the feasibility of a smartphone application-based dark adaptation (DA) measurement method (MOBILE-DA).
Methods:
On a Samsung Galaxy S8 smartphone, MOBILE-DA presented a 1.5° flashing stimulus (wavelength = 453 nm) between -1.15 and -4.33 log candela (cd)/m2 at 8° eccentricity using an adaptive staircase, and logged timing of user response (tapping on the screen) whenever the stimulus became visible (monocularly). In a dark room, the smartphone was placed ≈40 cm from the subject, and a white smartphone screen at maximum brightness (≈300 cd/m2) for 120 seconds was used for bleaching before testing. MOBILE-DA was evaluated in normally-sighted (NV) subjects (n = 15; age, 22-82 years). Additionally, a subject with myopic retinal degeneration (MRD; VA, 20/100; age, 62 years) and another with optic nerve atrophy (ONA; visual acuity [VA], 20/500; age, 40 years) were measured. Maximum test timing was capped at 20 minutes. Linear regression was performed to determine age-effect on DA parameters: rod-cone break time (t ) and test-time (t ). Use of the normalized area under the DA characteristics (AUC) as an outcome measure was explored.
Results:
For NV, the repeatability coefficients for t , t , and AUC were ±2.1 minutes, ±5.4 minutes, and 4.4%, respectively, and aging-related delays were observed (t , R 2 = 0.47, P = 0.003; t , R 2 = 0.34, P = 0.013; AUC, R 2 = 0.41, P = 0.006). Compared to ONA and NV, DA was greatly prolonged in the MRD subject (52% larger AUC than the NV mean).
Conclusion:
The age-effect was verified for MOBILE-DA measurements in NV subjects; impaired DA in a case with retinal-degeneration was observed.
Translational Relevance:
This study establishes feasibility of the smartphone-based DA measurement method as a potential accessible screening tool for various vision disorders.
More Related Videos
Related Concept Videos
Dark Triad and Person Perception
Adaptability of Cytoskeletal Filaments
Natural Selection and Adaptation
Beyond physical adaptations,...
Introduction to Innate and Adaptive Immunity
Innate immunity is the body's natural, nonspecific defense system that acts quickly to protect against pathogens. It incorporates physical barriers like skin and mucous membranes and cellular elements such as phagocytes and natural killer cells. This part of our immune system provides an immediate,...
Self-Evaluation: Self-Enhancement and Self-Verification
Non-ohmic Devices
Consider a simple circuit consisting of a battery, a diode, and a resistor. A...