Calibration-free beam hardening correction for myocardial perfusion imaging using CT

Jacob Levi1, Brendan L Eck2, Rachid Fahmi3

  • 1Department of Physics, Case Western Reserve University, Cleveland, OH, 44106, USA.

Medical Physics
|January 29, 2019
PubMed

Insights

A new automatic beam hardening correction (ABHC) method reduces artifacts in computed tomography myocardial perfusion imaging (CT-MPI). This calibration-free approach improves the accuracy of myocardial blood flow (MBF) calculations, enhancing CT's role in noninvasive coronary angiography.

Area of Science:

  • Medical Imaging
  • Radiology
  • Cardiovascular Imaging

Background:

  • Computed tomography myocardial perfusion imaging (CT-MPI) and coronary CT angiography (CTA) show promise for noninvasive assessment of coronary artery disease.
  • Beam hardening (BH) artifacts in conventional CT hinder accurate myocardial blood flow (MBF) quantification in CT-MPI.
  • Existing BH correction methods often require specialized hardware (e.g., energy-sensitive CT) or calibration procedures.

Purpose of the Study:

  • To propose and evaluate a novel calibration-free, automatic BH correction (ABHC) algorithm for CT-MPI.
  • To assess the efficacy of ABHC in reducing BH artifacts in static CT-MPI images.
  • To determine the impact of ABHC on the accuracy of dynamic CT-MPI-derived MBF measurements.

Main Methods:

  • Developed an iterative algorithm that optimizes polynomial BH correction parameters to minimize a BH-sensitive cost function.
  • Image segmentation identified soft tissue and highly attenuating materials (bones, iodine) for targeted correction.
  • Evaluated ABHC on simulated, phantom, preclinical porcine, and clinical CT-MPI datasets using conventional and spectral CT scanners.

Main Results:

  • ABHC significantly reduced BH streak artifacts and cupping in phantom studies (e.g., 86% reduction in cupping).
  • In preclinical CT-MPI, BH artifact was reduced from 24 ± 6 HU to < 5 ± 4 HU, with improved myocardial ROI uniformity.
  • ABHC reduced the coefficient of variation in MBF maps from 22% to 9%, approaching the 5% seen with virtual keV images.

Conclusions:

  • The developed automatic BH correction algorithm effectively reduces beam hardening artifacts in conventional CT.
  • ABHC improves the accuracy of CT-MPI by mitigating misinterpretation of reduced flow regions as actual deficits.
  • This method enhances CT-MPI's potential as a reliable noninvasive gatekeeper for invasive coronary angiography.
Abstract

Related Concept Videos

Imaging Studies I: CT and MRI01:14

Imaging Studies I: CT and MRI

Introduction: MRI and CT scans are crucial advancements in medical imaging techniques, playing a vital role in diagnosing conditions related to the gastrointestinal (GI) system. Each scan serves distinct purposes, targets specific areas, and requires unique nursing duties.
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
863
Imaging Studies for Cardiovascular System V: CT01:28

Imaging Studies for Cardiovascular System V: CT

Cardiac computed tomography (CT) scanning is an advanced cardiac imaging technique that utilizes CT technology, with or without intravenous (IV) contrast, to produce accurate cross-sectional virtual slices of specific areas of the heart, coronary circulation, and major blood vessels such as the aorta, pulmonary veins, and arteries. The computer processes these slices to generate three-dimensional images. Multidetector CT (MDCT) is a rapid form of CT scanning that captures multiple slices...
355
Glassware Calibration01:11

Glassware Calibration

Accurate calibration of glassware, such as volumetric flasks, pipettes, and burettes, is essential to ensure accurate measurements in the analytical laboratory. Calibration helps maintain consistency across measurements and prevents errors arising from inaccurate volumes.
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
1.4K
Distance Corrections01:15

Distance Corrections

To achieve precise distance measurements, especially in surveying and construction, certain corrections must be applied to account for potential sources of error like the standardization errors, temperature variations, and slope adjustments.Standardization error emerges when measurement equipment undergoes changes, such as wear, repairs, or weather impacts. To address this, surveyors compare the equipment’s readings to a standard. This process identifies any deviation that might lead to...
284
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT01:25

Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT

Calcium-Scoring CT ScanA calcium-scoring CT scan, also known as coronary artery calcium (CAC) scan, detects calcium deposits in the coronary arteries. This test assesses the risk of coronary artery disease (CAD), which can lead to cardiovascular events such as angina, heart failure, and sudden cardiac arrest.A calcium-scoring CT scan is generally recommended for individuals at intermediate risk of CAD without symptoms. It includes:Men aged 40-75 and women aged 50-75: Especially those with a...
476
Beams01:30

Beams

Beams are integral components of structural engineering and construction, designed to support loads applied at various points along their length. These long, straight members can be classified based on geometry, cross-section, support type, and equilibrium condition.
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
1.8K