Technical Note: Relative proton stopping power estimation from virtual monoenergetic images reconstructed from
Guillaume Landry1, Fabian Dörringer1, Salim Si-Mohamed2,3
1Department of Medical Physics, Faculty of Physics, Ludwig-Maximilians-Universität München, Munich, Germany.
Purpose:
The objective of this technical note was to investigate the accuracy of proton stopping power relative to water (RSP) estimation using a novel dual-layer, dual-energy computed tomography (DL-DECT) scanner for potential use in proton therapy planning. DL-DECT allows dual-energy reconstruction from scans acquired at a single x-ray tube voltage V by using two-layered detectors.
Methods:
Sets of calibration and evaluation inserts were scanned at a DL-DECT scanner in a custom phantom with variable diameter D (0 to 150 mm) at V of 120 and 140 kV. Inserts were additionally scanned at a synchrotron computed tomography facility to obtain comparative linear attenuation coefficients for energies from 50 to 100 keV, and reference RSP was obtained using a carbon ion beam and variable water column. DL-DECT monoenergetic (mono-E) reconstructions were employed to obtain RSP by adapting the Yang-Saito-Landry (YSL) method. The method was compared to reference RSP via the root mean square error (RMSE) over insert mean values obtained from volumetric regions of interest. The accuracy of intermediate quantities such as the relative electron density (RED), effective atomic number (EAN), and the mono-E was additionally evaluated.
Results:
The lung inserts showed higher errors for all quantities and we report RMSE excluding them. RMSE for μ from DL-DECT mono-E was below 1.9%. For the evaluation inserts at D = 150 mm and V = 140 kV, RED RMSE was 1.0%, while for EAN it was 2.9%. RSP RMSE was below 0.8% for all D and V, which did not strongly affect the results.
Conclusions:
In this investigation of RSP accuracy from DL-DECT, we have shown that RMSE below 1% can be achieved. It was possible to adapt the YSL method for DL-DECT and intermediate quantities RED and EAN had comparable accuracy to previous publications.
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Virtual Work
In static equilibrium, a body can experience an imaginary or virtual movement, such as displacement or rotation. The virtual work done by a force is equal to the dot product of force and virtual displacement in the direction of the force. When it comes to virtually rotating a...
What are Estimates?
The estimate for the mean of a sample is denoted by ͞x, whereas the mean of the population is designated as μ. Further, parameters such...
Power