Related Experiment Video
Updated: Jan 30, 2026

Pilot In Vitro Study to Assess Cleaning Ability and Effects of Different Decontamination Methods on Implant Surfaces
Published on: November 21, 2025
Effects of corrosion and cleaning method on taper dimensions: an in vitro investigation
Ahmad Faizan1, Jason Longaray1, Jim E Nevelos1
1Stryker, Mahwah, NJ, USA.
Purpose:
Taper corrosion related revisions have recently been reported in the orthopaedic literature. Cleaning procedure of the trunnions during hip revision is not standardised. The purpose of this bench top investigation was to understand the alterations in the trunnion dimensions and surface roughness characteristics as a result of corrosion product build-up.
Methods:
8 titanium alloy trunnions and CoCr femoral heads assemblies were cyclically tested in a mechanical simulator. Following disassembly of the tested constructs, the trunnions were cleaned using 2 methods. The trunnion dimensions were measured using coordinate measuring machine, and surface roughness was measured using white light interferometry. The trunnions were reassembled with ceramic femoral heads and titanium sleeves following cleaning. Head/sleeve pull-off testing was conducted to understand the effects of cleaning methods on the pull-off strength.
Results:
Grade 4 corrosion was observed on all trunnions after mechanical testing. The aggressive cleaning methods had a larger impact on the surface roughness when compared to the light cleaning method. The aggressive cleaning method also decreased the taper cone angle. The pull-off strength was not affected by the cleaning method and the pull-off values were approximately 50% of the assembly loads.
Conclusions:
The study suggests that trunnion cleaning method may alter the surface roughness and taper cone angle of the existing trunnion. However, the effects of these changes on the pull-off strength did not reach statistical significance. Complex corrosion testing under cyclic loading conditions are warranted to understand the long-term effects of these changes.
Related Concept Videos
Corrosion
Corrosion of Reinforcement
However, over time and under certain conditions like carbonation, chloride ingress, and cracking this protective state can be compromised. Steel has areas with...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Support Reactions in Three Dimensions
Ball and Socket Joint is one of the supports allowing free rotation about any axis. This freedom of rotation is...
Relative Velocity in One Dimension
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...