Safety and Image Quality of 1.5-T Endorectal Coil Multiparametric MRI of the Prostate or Prostatectomy Fossa for
Takashi Tanaka1,2, Adam T Froemming3, Anshuman Panda1
11 Department of Radiology, Mayo Clinic, 13400 E Shea Blvd, Scottsdale, AZ 85259.
Objective:
The purpose of this study is to report the patient safety and image quality of 1.5-T multiparametric MRI of the prostate in patients with cardiac implantable electronic devices (CIEDs).
Materials And Methods:
In this retrospective study, a database was searched to identify prostate multiparametric 1.5-T MRI examinations performed with endorectal coils for patients with CIEDs from 2012 to 2016 (study group) and matched patients without CIEDs (control group). Clinical safety in the study group was reviewed. The specific absorption rate (SAR) and signal-to-noise ratio (SNR) were measured in both groups. Imaging quality and artifact on T2-weighted images, DW images, and dynamic contrast-enhanced images were rated on a 5-point scale by two independent readers.
Results:
The study group consisted of total 28 multiparametric MRI examinations in 25 patients. There were no serious device-related adverse effects observed (0/28; 0%), and the estimated whole-body SAR in the study group was never greater than 1.5 W/kg. The SNR values tended to be lower in the study group than in the control group. However, overall perceived image preferences and influences of artifacts on image quality for the study group were not significantly different from those for the control group (p > 0.05), which were rated above average (rating 3) by both readers 1 and 2.
Conclusion:
Multiparametric 1.5-T MRI examination of the prostate can be safely performed in selected patients with CIEDs under controlled conditions with applicable image quality while maintaining a SAR less than 1.5 W/kg.
Insights
Prostate multiparametric 1.5-T MRI is safe for patients with cardiac implantable electronic devices (CIEDs). Image quality is maintained, with no serious adverse effects and acceptable signal-to-noise ratio.
Area of Science:
- Radiology
- Medical Imaging
- Cardiovascular Devices
Background:
- Cardiac implantable electronic devices (CIEDs) can pose safety and image quality challenges during MRI.
- Multiparametric MRI (mpMRI) is crucial for prostate cancer diagnosis and management.
Purpose of the Study:
- To evaluate the patient safety and diagnostic image quality of 1.5-T prostate mpMRI in patients with CIEDs.
Main Methods:
- Retrospective analysis of 1.5-T prostate mpMRI in patients with CIEDs (study group) and without (control group).
- Assessed clinical safety, specific absorption rate (SAR), signal-to-noise ratio (SNR), and image quality/artifacts.
- Two independent readers rated image quality on a 5-point scale.
Main Results:
- No serious device-related adverse events occurred in the CIED group (0/28 examinations).
- Whole-body SAR remained below 1.5 W/kg.
- While SNR was slightly lower in the CIED group, image quality and artifact perception were not significantly different from controls (rated above average).
Conclusions:
- 1.5-T prostate mpMRI can be safely performed in selected patients with CIEDs under controlled conditions.
- Applicable image quality is achievable, maintaining safety and diagnostic utility.
- Adherence to SAR limits (<1.5 W/kg) is essential for safe scanning.
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Survey Safety
Household Wiring And Electrical Safety
Quality Control
Quality control helps track data, visualize trends, and identify variations, making it easier to detect deviations that may affect the accuracy of an analysis. One way to do this is by generating a quality control chart, which...
Quality Assurance
Quality of Water