Image-plate sensitivity to x rays at 2 to 60 keV
M J Rosenberg1, D B Thorn2, N Izumi2
1Laboratory for Laser Energetics, University of Rochester, Rochester, New York 14623, USA.
Researchers measured the sensitivity of Fuji image plates (IPs) for X-ray spectrometers. Sensitivity varied by up to 5% with scanner settings, highlighting the need for cross-calibration between facilities.
Area of Science:
- Nuclear Physics
- Materials Science
- Spectroscopy
Background:
- Fuji SR and MS image plates (IPs) are crucial detectors in X-ray spectrometers.
- Accurate sensitivity calibration is essential for quantitative analysis in high-energy physics experiments.
Purpose of the Study:
- To quantify the sensitivity of Fuji SR and MS image plates (IPs) across a broad X-ray energy range.
- To investigate the impact of scanner settings and pixel size on IP sensitivity.
- To assess the fading properties of MS-type image plates.
Main Methods:
- Measured IP sensitivity using radioisotopes (6-60 keV) and a Manson source (1.5-8 keV).
- Systematically examined IP sensitivity variations with scanner settings and pixel size.
- Evaluated MS-type IP fading over extended periods.
Main Results:
- IP sensitivity showed variations of approximately 5% for specific types and scanner settings.
- IP sensitivity varied by up to a factor of 2 depending on IP type and scanner parameters.
- Radioisotope technique provided insights into MS-type IP fading characteristics.
Conclusions:
- Cross-calibration of image plate scanners across different facilities is essential.
- A consistent absolute sensitivity curve from 2-60 keV requires inter-facility calibration.
- Understanding IP sensitivity variations is critical for accurate X-ray spectroscopy data.
More Related Videos
Related Concept Videos
X-ray Imaging
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
X-ray Diffraction of Biological Samples
According to Bragg's law, when X-rays strike the sample positioned on a stage, the rays are scattered by the electron clouds around the sample atoms. The X-ray diffraction or scattering is caused by constructive interference of the X-ray waves that reflect off the internal...
meta-Directing Deactivators: –NO2, –CN, –CHO, –⁠CO2R, –COR, –CO2H
Radiological Investigation I: X-ray and CT