Influence of iterative reconstruction and dose levels on metallic artifact reduction: A phantom study within four CT
J Greffier1, A Larbi1, J Frandon1
1Department of Radiology, CHU Nîmes, Université Montpellier, Medical Imaging Group Nîmes, EA 2415, 30029 Nîmes cedex, France.
Purpose:
To compare metallic artifact reduction (MAR) algorithms proposed by four vendors according to the delivered dose and iterative level using a phantom study.
Methods:
Four CT systems (Revolution GSI®, Ingenuity Elite®, Somatom Edge®, and Aquilion Prime®) equipped with MAR algorithms (Smart MAR®, O-MAR®, iMAR®, and SEMAR®) were compared. Acquisitions were performed with CIRS model 062M Phantom containing a titanium rod core insert using 120kV and two dose levels (3 and 7mGy). Images were reconstructed with and without MAR algorithms using standard "soft tissue" kernel for filtered back projection (FBP) and intermediary iterative level. Artifact propagation was assessed by counting the number of pixels containing a HU outside a defined threshold interval (>100HU and<-80 HU). Artifact correction was evaluated by computing the differences between images with and without MAR.
Results:
Accuracy of NCT values increased significantly using MAR algorithms, IR, and high dose levels (P<0.001). Image noise reduced -31±15 (SD) % (range: -50%;-14%) with Smart-MAR®, -28±3 (SD) % (range -31%; -25%) with O-MAR®, -32±7 (SD) % (range: -40%; -24%) with iMARCN®, -52±8 (SD) % (range: -60%; -42%) with iMARTH® and -29±6 (SD) % (range: -37%; -23%) with SEMAR®. The number of pixels outside the threshold interval was also reduced using MAR algorithms. Each MAR algorithm corrected in distinct patterns, with satisfactory artifact correction for all MAR algorithms.
Conclusion:
This study demonstrates that artifact correction using MAR algorithms differs according to the main manufacturers, although corrections are satisfactory for all systems. Corrections also improved by using IR and increasing the dose level.
More Related Videos
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
12:38Structure of HIV-1 Capsid Assemblies by Cryo-electron Microscopy and Iterative Helical Real-space Reconstruction
Published on: August 9, 2011
Related Concept Videos
Influence of Earth's Curvature and Atmospheric Refraction on Leveling
Metal-Ligand Bonds
In these complexes, transition metals form coordinate covalent bonds, a kind of Lewis acid-base interaction in which both of the electrons in the bond are contributed by a donor (Lewis base) to an electron acceptor (Lewis acid). The Lewis acid in...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Oxidation-Reduction Reactions
Bioavailability Study Design: Single Versus Multiple Dose Studies