Related Experiment Video
Updated: Jan 29, 2026

Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025
A biomechanical comparison of new techniques for distal clavicular fracture repair versus locked plating
Gautam P Yagnik1, Paul C Brady2, Joseph P Zimmerman1
1Miami Orthopaedics and Sports Medicine Institute, Baptist Health South Florida, Coral Gables, FL, USA.
Background:
Unstable distal clavicular fractures treated surgically are associated with high failure rates and hardware-related complications. Newer techniques have shown promising early clinical results with fewer hardware complications; however, their biomechanical performance has not been assessed. This study biomechanically compared a distal-third locking plate with 3 newer techniques that incorporate coracoid fixation into the construct.
Methods:
The study randomized 36 adult fresh frozen cadaveric shoulders to 4 groups: (1) distal-third locking plate (P); (2) distal-third locking plate with a coracoid button augmentation (P + CB); (3) coracoclavicular button (CB); and (4) coracoclavicular button with coracoclavicular ligament reconstruction using semitendinosus allograft (CB + CC). After fixation, each specimen was stressed in the coronal plane. Cyclic displacement, load at 10-mm displacement, and ultimate load to failure were measured.
Results:
All 3 experimental groups biomechanically outperformed the locking plate. Mean load to failure was significantly higher in the CB (343 ± 76 N) and CB + CC (349 ± 94 N) groups compared with the P group (193 ± 52 N). There was also significantly less cyclic displacement in the CB (4.3 ± 1.9 mm) and CB + CC (4.4 ± 1.9 mm) groups compared with the P group (8.2 ± 2.9 mm). With respect to load at 10 mm of displacement, which essentially measures a clinical failure, the P + CB (235 ± 112 N), CB (253 ± 111 N), and CB+CC (238 ± 76 N) experimental groups significantly outperformed the P group (96 ± 29 N).
Conclusions:
CB and CB + CC techniques demonstrated more than 75% greater strength than the traditional locking plate alone. Coupled with greater overall construct strength and lower-profile hardware, these newer techniques may result in improved clinical outcome and fewer hardware-related complications.
More Related Videos
15:11Surgical Fixation of Sternal Fractures: Preoperative Planning and a Safe Surgical Technique Using Locked Titanium Plates and Depth Limited Drilling
Published on: January 5, 2015
06:41A Minimally Invasive Model to Analyze Endochondral Fracture Healing in Mice Under Standardized Biomechanical Conditions
Published on: March 22, 2018
Related Concept Videos
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the...
Mismatch Repair
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Overview of DNA Repair
Chemically...
Base Excision Repair
The first step of...
Nucleotide Excision Repair