Reproducibility of BiliCare™ Transcutaneus Bilirrubin Meter in Mexican Newborns
David Oldak1, Gabriela García1, Estela E Gonzalez1
1Division of Neonatology, Department of Pediatrics, Angeles de las Lomas Hospital, State of México, Mexico.
Background:
Newborn hyperbilirubinemia is considered a worldwide health problem that demands medical evaluation. Noninvasive transcutaneous bilirubin (TcB) has been used as a screening method with different devices but there has not been any evaluation of reproducibility of the same brand devices. The BiliCare™ system is evaluated to demonstrate consistency between measurements with four different devices.
Methods:
336 TcB measurements were obtained with four BiliCare™ devices in 21 Mexican icteric newborns with a mean postnatal age of 44.1 hours of life and 38 weeks of gestation (33-41). Two measurements were taken in the same ear alternatively at the scaphoid fossa with each device. TcB values were compared between devices. Validity was compared with total serum bilirubin (TB).
Results:
intraclass correlation coefficient demonstrates a minimum limit in the study of 0.945 and maximum of 0.988 with the same device. Correlations with serum and between devices gave results above 0.932.
Conclusions:
BiliCare™ transcutaneous bilirubin measurement instrument has very good intra- and interdevice reproducibility; also correlation of TcS with serum bilirubin gave statistically the same results.