Evaluation of a standardized protocol for thrombin generation using the calibrated automated thrombogram: A Nordic
Marcus Ljungkvist1, Karin Strandberg2, Erik Berntorp1
1Department of Translational Medicine & Centre for Thrombosis and Haemostasis, Lund University, Malmö, Sweden.
Standardizing the thrombin generation assay-calibrated automated thrombogram (TGA-CAT) improves results. Two reference plasmas, HemosIL RP and Affinity RP, effectively normalized TGA-CAT data, enhancing its reliability for clinical use.
Area of Science:
- Coagulation science
- Clinical chemistry
- Laboratory diagnostics
Background:
- The thrombin generation assay-calibrated automated thrombogram (TGA-CAT) measures overall plasma coagulation capacity.
- Current lack of standardization limits TGA-CAT's use as a research tool.
Purpose of the Study:
- To enhance TGA-CAT standardization through a detailed protocol.
- To evaluate the efficacy of three reference plasmas (RPs) in normalizing TGA-CAT results.
Main Methods:
- Six Nordic centers collaborated to establish a detailed laboratory standardization protocol for TGA-CAT.
- Hypo-, normal-, and hypercoagulable plasmas were analyzed.
- Three commercial lyophilized RPs were used for data normalization.
Main Results:
- Before normalization, inter-laboratory variability was generally low (≤10%), with exceptions for endogenous thrombin potential and peak in hypocoagulable plasma.
- Normalization using HemosIL RP and Affinity RP significantly improved result variability.
- One RP did not effectively normalize results.
Conclusions:
- The developed standardization concept yielded TGA-CAT results comparable to routine coagulation assays.
- HemosIL RP is suitable for normalizing samples from populations with low or unknown coagulability.
- Affinity RP is recommended for analyzing plasma samples from populations with known or suspected hypercoagulability.
Related Concept Videos
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Calibration Curves: Correlation Coefficient
Plotting and Calibrating the Root Locus
The maximum gain occurs at the breakaway points between open-loop poles on the real axis, while the minimum gain is...
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...
Standard Enthalpy of Formation