Bacterial Contamination of Circuit Inner Surfaces After High-Flow Oxygen Therapy
Mutsuo Onodera1, Emiko Nakataki1, Nobuto Nakanishi1
1Emergency and Critical Care Medicine, Tokushima University Graduate School, Tokushima, Japan.
Background:
During high-flow oxygen therapy, heated humidified gas is delivered at high flow. Although the warmth and humidity of this gas facilitates mucociliary function in the lower airway, warm and humid conditions also promote bacterial growth. Bacterial contamination of high-flow oxygen therapy circuits has not been well investigated. We examined the incidence of bacterial contamination in high-flow oxygen therapy circuits.
Methods:
This was a prospective observational study in a university hospital 10-bed general ICU. After final high-flow oxygen therapy discontinuation, samples were obtained from the interface and the chamber ends of the circuit. Initially, a half circumference of each inner surface was swabbed, after which the whole circuit was left in the ICU at room temperature for 6 h. The samples were then, in the same way, taken from the previously unswabbed arcs of the end inner surfaces. All the samples were sent to a biology laboratory and cultured.
Results:
In all, 118 samples were collected from 31 circuits. The median duration of high-flow oxygen therapy was 48 h (interquartile range, 26-96 h). Of 31 circuits, contamination occurred in 5 circuits (16.1% [95% CI 5.5-33.7%]).
Conclusions:
Bacterial contamination of inner surfaces of the circuit after high-flow oxygen therapy was relatively high.
Related Concept Videos
Contaminants and Errors
Another key consideration is determining the appropriate number of samples required to...
Treatment for Pulmonary Arterial Hypertension: Oxygen Therapy for Respiratory Failure
Oxygen therapy is vital in increasing and maintaining blood oxygen levels in PAH patients. As a result, it aids in reducing fatigue,...
Second-Order Circuits
Input signals typically originate from voltage or current sources, with the output often representing voltage across the capacitor and/or current through the inductor. For example, in...
Gene Therapy
Bacterial Signaling
First-Order Circuits
One common example of a first-order circuit is the RC (resistor-capacitor) circuit. These circuits are used in relaxation oscillators such as neon lamp oscillator circuits. When voltage is...