Biomechanical evaluation of location and mode of failure in three screw fixations for a comminuted transforaminal
Brett D Crist1, Ferris M Pfeiffer1,2, Michael S Khazzam3
1Department of Orthopaedic Surgery, University of Missouri, Columbia, 1100 Virginia Ave., Columbia, MO 65212, USA.
Background:
Pelvic ring-comminuted transforaminal sacral fracture injuries are rotationally and vertically unstable and have a high rate of failure.
Objective:
Our study purpose was to use three-dimensional (3D) optical tracking to detect onset location of bone-implant interface failure and measure the distances and angles between screws and line of applied force for correlation to strength of pelvic fracture fixation techniques.
Methods:
3D relative motion across sacral-rami fractures and screws relative to bone was measured with an optical tracking system. Synthetic pelves were used. Comminuted transforaminal sacral-rami fractures were modelled. Each pelvis was stabilised by either (1) two iliosacral screws in S1, (2) one transsacral screw in S1 and one iliosacral screw in S1 and (3) one trans-alar screw in S1 and one iliosacral screw in S1; groups 4-6 consisted of fixation groups with addition of anterior inferior iliac pelvic external fixator. Eighteen-instrumented pelvic models with right ilium fixed simulate single-leg stance. Load was applied to centre of S1 superior endplate. Five cycles of torque was initially applied, sequentially increased until permanent deformation occurred. Five cycles of axial load compression was next applied, sequentially increased until permanent deformation occurred, followed by axial loading to catastrophic failure. A Student t test was used to determine significance (p < 0.05).
Results:
The model, protocol and 3D optical system have the ability to locate how sub-catastrophic failures initiate. Our results indicate failure of all screw-based constructs is due to localised bone failure (screw pull-in push-out at the ipsilateral ilium-screw interface, not in sacrum); thus, no difference was observed when not supplemented with external fixation.
Conclusion:
Inclusion of external fixation improved resistance only to torsional loading.
Translational Potential Of This Article:
Patients with comminuted transforaminal sacral-ipsilateral rami fractures benefit from this fixation.
More Related Videos
Related Concept Videos
Heart Failure IV: Classification and Diagnostic Evaluation
Screw: Problem Solving
To evaluate the...
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
What is a Mode?
There can be more than one mode in a data set if multiple values have the same highest frequency. For instance, suppose that the Statistics exam scores of 20 students are: 50; 53; 59; 59; 63; 63; 72; 72; 72; 72; 72; 76; 78; 81; 83; 84; 84; 84; 90; 93. Here, the mode is 72, as it occurs most frequently, five times.
A data set with two modes is called bimodal. For example,...
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...