Optimizing radiation dose parameters in MDCT arthrography of the shoulder: illustration of basic concepts in a
Julien Aguet1, Fabio Becce1, Vincent Dunet1
1Department of Diagnostic and Interventional Radiology, Lausanne University Hospital and University of Lausanne, Rue du Bugnon 46, 1011, Lausanne, Switzerland.
Objective:
To determine in a cadaveric study the lowest achievable radiation dose and optimal tube potential generating diagnostic image quality in multidetector computed tomography (MDCT) arthrography of the shoulder.
Materials And Methods:
Six shoulders from three human cadavers were scanned using a 256-MDCT system after intra-articular injection of diluted iodinated contrast material. Using six decreasing radiation dose levels (CTDIvol: 20, 15, 10, 8, 6, and 4 mGy) and for each dose level, four decreasing tube potentials (140, 120, 100, and 80 kVp), image noise and contrast-to-noise ratio (CNR) were measured. Two independent and blinded observers assessed the overall diagnostic image quality, subjective amount of noise, and severity of artifacts according to a four-point scale. Influence of those MDCT data acquisition parameters on objective and subjective image quality was analyzed using the Kruskal-Wallis and Wilcoxon signed-rank tests, and pairwise comparisons were performed.
Results:
Multidetector CT protocols with radiation doses of 15 mGy or higher, combined with tube potentials of 100 kVp or higher, were equivalent in CNR to the reference 20 mGy-140 kVp protocol (all p ≥ 0.054). Above a CTDIvol of 10 mGy and a tube potential of 120 kVp, all protocols generated diagnostic image quality and subjective noise equivalent to the 20 mGy-140 kVp protocol (all p ≥ 0.22).
Conclusions:
Diagnostic image quality in MDCT arthrography of the shoulder can be obtained with a radiation dose of 10 mGy at an optimal tube potential of 120 kVp, corresponding to a reduction of up to 50% compared with standard-dose protocols, and as high as 500% compared with reported protocols in the literature.
More Related Videos
Related Concept Videos
Basic Concept
Length, which measures the distance traveled by an object, is a fundamental concept in engineering mechanics. We use coordinates relative to a reference point to describe the distance. Length not only helps to describe the...
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
Biological Effects of Radiation
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations
Muscles of the Shoulder
Anterior Thoracic Muscles
The anterior thoracic muscles include the serratus anterior, subclavius, and...
Bioavailability Study Design: Single Versus Multiple Dose Studies