Evidence From a Long-Term, Systematic Post-Market Surveillance Program: Clinical Performance of a
Steven Setford1, Stuart Phillips1, Mike Grady1
1LifeScan Scotland, Inverness, UK.
Background:
Described is a manufacturer's systematic post-market evaluation of the long-term clinical accuracy of a commercially available blood glucose monitoring (BGM) test strip product.
Methods:
Production batches of test strips were routinely and regularly sampled and evaluated in a clinical setting to assess product accuracy. Evaluations were performed on capillary blood samples from a minimum of 100 subjects with diabetes, by clinical staff according to instructions for use. Readings were compared against capillary blood samples collected at the same time and measured by a standard laboratory reference method. Clinical accuracy was calculated according to EN ISO 15197:2015.
Results:
A total of 21 115 paired results were obtained, equating to 209 production batches over the >3-year period since test strip launch. Of the results, 97.6% met the accuracy criterion (range: 97.1-98.1% by year), with 98.1% of values presenting zero risk as defined by the surveillance error grid. At the <5th (21.0-33.8%) and >95th (48.3-59.4%) percentile extremes of hematocrit distribution, 97.9% and 96.4% of values were clinically accurate. The product also demonstrated clinical accuracy across all seven glucose ranges ("bins") as defined by the standard. Under conditions of combined hematocrit and glucose (<80 mg/dL and ≥300 mg/dL) extremes, 97.7% of values were clinically accurate.
Conclusions:
Methodologies and results from a manufacturer's self-imposed clinical accuracy surveillance program of a BGM product is presented. Given the publication of sometimes-conflicting data presented within ad hoc BGM clinical accuracy evaluations, usually of limited size, it is advocated that BGM manufacturers adopt similarly robust and systematic surveillance programs to safeguard patients.
Related Concept Videos
Glucose Homeostasis: Regulation of Blood Glucose
During fasting, when blood glucose levels are low, the pancreas secretes glucagon. it...
Hormones Regulating Blood Glucose
In addition to accelerating glucose uptake and utilization, insulin has...
The Evidence for Evolution
Voltammetry: Stripping Methods
Anodic Stripping Voltammetry (ASV)
ASV is used to determine metals and metalloids at trace levels. It involves two steps: deposition and stripping. First, a negative potential is applied to the...
Principles of Disease Surveillance
Insensitive Nuclei Enhanced by Polarization Transfer (INEPT)