Performance and Usability of Three Systems for Continuous Glucose Monitoring in Direct Comparison
Guido Freckmann1, Manuela Link1, Ulrike Kamecke1
11 Institut für Diabetes-Technologie, Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm, Ulm, Germany.
Background:
To be able to compare continuous glucose monitoring (CGM) systems, they have to be worn in parallel by the same subjects. This study evaluated the performance and usability of three different CGM systems in direct comparison.
Method:
In this open, prospective study at two sites, 54 patients with diabetes wore three CGM systems each (Dexcom G5™ Mobile CGM system [DG5], Guardian™ Connect system [GC], and a Roche CGM system [RCGM]) in parallel for 6 or 7 days in a mixed inpatient and outpatient setting. Capillary comparison measurements were performed using a self-monitoring of blood glucose (SMBG) system. During study site visits, glucose excursions were induced. Performance of the systems was evaluated by calculating mean absolute relative differences (MARD, calculated as absolute differences for glucose concentrations <100 mg/dL and as relative differences for glucose concentrations ≥100 mg/dL), and mean relative differences (MRD, bias) between CGM and SMBG results. In addition, usability of the systems was assessed.
Results:
Overall MARD was 10.1 ± 2.1 for DG5, 11.5 ± 4.2 for GC, and 11.9 ± 5.6 for RCGM. Performance improved in all systems after the first day of use. All systems showed >99% of values within zones A and B of the consensus error grid. Overall, all CGM systems showed a small negative bias compared to SMBG. Usability of the systems differed regarding patch adhesion rate, failure rate, and patient rating. Most patients preferred GC, but in general all systems were rated positively.
Conclusion:
All three CGM systems showed similar overall accuracy in this direct comparison, but small differences were observed with regard to specific glucose ranges and usability aspects.
Related Concept Videos
BIBO stability of continuous and discrete -time systems
To determine the BIBO stability, the convolution integral is utilized when a bounded continuous-time input is applied to a Linear Time-Invariant (LTI) system....
Glucose Homeostasis: Regulation of Blood Glucose
During fasting, when blood glucose levels are low, the pancreas secretes glucagon. it...
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Glucose Transporters
Facilitated diffusion-glucose transporters (GLUTs) are encoded by the solute-linked carrier (SLC) family 2, subfamily A gene family, or SLC2A. The 14 GLUT protein members are distributed into three classes:
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Continuing Care