A Comparison between 6-point Dixon MRI and MR Spectroscopy to Quantify Muscle Fat in the Thigh of Subjects with
1Alexandra Grimm, Institute of Medical Physics, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Henkestr. 91, 91052 Erlangen, Germany, E-mail address: alexandra.grimm@fau.de, Fax number: +49 9131 8522824, Telephone number: +49 9131 8525535.
Background:
Changes in muscle fat composition as for example observed in sarcopenia, affect physical performance and muscular function, like strength and power.
Objectives:
The purpose of this study was to compare 6-point Dixon magnetic resonance imaging and multi-echo magnetic resonance spectroscopy sequences to quantify muscle fat. Setting, participants and measurements: Two groups were recruited (G1: 23 healthy young men (28 ± 4 years), G2: 56 men with sarcopenia (80 ± 5 years)). Proton density fat fraction was measured with a 6-point product and a 6-point prototype Dixon sequence in the left thigh muscle and with a high-speed multi-echo T2*-corrected H1 magnetic resonance spectroscopy sequence within the semitendinosus muscle of the left thigh. To evaluate the comparability among the different methods, Bland-Altman and linear regression analyses of the proton density fat fraction results were performed.
Results:
Mean differences ± 1.96 * standard deviation between spectroscopy and 6pt Dixon sequences were 1.9 ± 3.3% and 1.5 ± 3.6% for the product and prototype sequences, respectively. High correlations were measured between the proton density fat fraction results of the 6-point Dixon sequences and spectroscopy (R = 0.95 for the product sequence and R = 0.97 for the prototype sequence).
Conclusions:
Dixon imaging and spectroscopy sequences show comparable accuracy for fat measurements in the thigh. Spectroscopy is a local measurement, whereas Dixon sequences provide maps of the fat distribution. The high correlations of the 6-point Dixon sequences with spectroscopy support their clinical use. They provide higher spatial resolution than spectroscopy, but are not suitable for a more complicated spectral analysis to separate extra- and intramyocellular lipids.
More Related Videos
09:40Phosphorus-31 Magnetic Resonance Spectroscopy: A Tool for Measuring In Vivo Mitochondrial Oxidative Phosphorylation Capacity in Human Skeletal Muscle
Published on: January 19, 2017
09:19A Guide to Examining Intramuscular Fat Formation and its Cellular Origin in Skeletal Muscle
Published on: May 26, 2022
Related Concept Videos
Muscles that Move the Thigh
Three other significant muscles are the gluteus maximus, gluteus medius, and gluteus minimus. The gluteus maximus originates from the posterior surface of the ilium, sacrum, and coccyx, and the thoracolumbar...
Peripherally and Centrally Acting Muscle Relaxants: A Comparison
Centrally acting muscle relaxants can be further divided into spasmolytic and antispasmodic drugs. Spasmolytic...
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Quantifying Work
Cable Subjected to Its Own Weight
A generalized loading function is employed to analyze a cable subjected to its own weight. This function considers the force acting along the cable's arc length rather than its projected length, providing a more accurate...
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...