Measuring liver fat fraction with complex-based chemical shift MRI: the effect of simplified sampling protocols on
Alexander J Procter1, Julia Y Sun2, Paul N Malcolm2
1Norfolk and Norwich University Hospital NHS Foundation Trust, Colney Ln, Norwich, NR4 7UY, UK. ajprocterac@gmail.com.
Background:
The assessment of liver percentage fat fraction (%FF) using proton density fat fraction sequences is becoming increasingly accessible. Previous studies have tended to use multiple small ROIs that focus on Couinaud segments. In an effort to simplify day-to-day analysis, this study assesses the impact of using larger, elliptical ROIs focused on a single hepatic lobe. Additionally, we assess the impact of sampling fewer transhepatic slices when measuring %FF.
Methods:
Retrospective analysis of prospectively obtained images from 34 volunteers using an IDEAL IQ sequence. Two observers independently measured %FF using three different protocols: freehand whole-liver ROI (fh-ROI), elliptical-ROI on the right lobe (rt-ROI) and elliptical-ROI on the left lobe (lt-ROI).
Results:
Inter-observer reliability for all measurements techniques was 'excellent' (Spearman's rank correlation coefficients 0.81-0.98). There was a significant difference (Paired Wilcoxon Test: p < 0.001) between the median %FF obtained using fh-ROI when compared to the rt-ROI method, the maximum mean difference between the two techniques was 2.79% (95% CI). For all sampling methods a Kruskall-Wallis analysis demonstrated no significant difference in mean %FF when the number of slices sampled was reduced from 11 to 1. The mean coefficient of variance increased when more slices were sampled (3 slices = 0.1, 11 slices = 0.17, p < 0.001).
Conclusion:
Simplified ROIs focused on one hepatic lobe provide %FF measurements that are unlikely to be sufficiently accurate for use in clinical practice. Freehand whole-liver ROIs should be used in preference. A single freehand ROI measurement taken at the level of the hepatic hilum yields a %FF that is representative of the mean whole liver % FF. Multiple slices are needed to measure heterogeneity.
Related Concept Videos
Uncertainty in Measurement: Accuracy and Precision
Proton (¹H) NMR: Chemical Shift
Absorption signals of all the protium nuclei...
NMR Spectroscopy: Chemical Shift Overview
For instance, the proton...
Inductive Effects on Chemical Shift: Overview
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...
π Electron Effects on Chemical Shift: Overview