Does thread shape affect the fixation strength of the bioabsorbable interference screws for anterior cruciate
Gerardo L Garcés1, Oscar Martel2, Alejandro Yánez2
1Department of Medical and Surgical Sciences, University of Las Palmas de Gran Canaria, Edificio de Ciencias de la Salud, Campus Universitario de San Cristobal, Trasera del Hospital InsularC/ Doctor Pasteur s/n, 35016, Las Palmas, Spain. ggarces@imqc.es.
Background:
The purpose of this study was to compare the biomechanical behaviour of two bioabsorbable interference screws with different geometries.
Methods:
Two different pitch (2.5 and 5 mm) bioabsorbable interference screws, both 9 × 30 mm, were tested. Tests were performed with forty bovine digital extensor tendons and skeletally mature porcine tibiae. Two protocols of cyclic tests at 1 Hz were performed: 1000 cycles from 50 to 250 N, and 5000 cycles from 100 to 300 N (n = 10 for each type of test and screw). After the cyclic loading, a final ramp displacement until failure at 0.5 mm/s was applied.
Results:
The stiffness after the cyclic phase of the tests was not statistically different between the two screws (1000th cycle: 2.5 mm pitch 280.3 ± 56.4 N/mm, 5 mm pitch 275.2 ± 65.0 N/mm, P = .965; 5000th cycle: 2.5 mm pitch 281.3 ± 66.4 N/mm, 5 mm pitch 286.1 ± 79.4 N/mm, P = .814). The yield load was not significantly different between the screws (1000 cycle tests: 2.5 mm pitch 482.2 ± 120.2 N, 5 mm pitch 495.9 ± 131.3 N, P = .508; 5000 cycle tests: 2.5 mm pitch 476.4 ± 65.3 N, 5 mm pitch 494.3 ± 39.2 N, P = .391). No correlation was found between the insertion torque and yield load (1000 cycle tests, R2 = 0.013; 5000 cycle tests, R2 = 0.006).
Conclusions:
The pitch of bioabsorbable interference screws does not seem to affect fixation strength. Also, the authors recommend not to use insertion torque alone to estimate the fixation strength.
Related Concept Videos
Interference and Diffraction
RNA Interference
This process occurs naturally in cells, often through the activity of genomically-encoded microRNAs. Researchers can take advantage of this mechanism by introducing synthetic RNAs to deactivate specific genes for research or therapeutic purposes. For example, RNAi could be used...
Screw: Problem Solving
To evaluate the...
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Role of Affect in Interpersonal Attraction
Acid Strength and Molecular Structure
In the absence of any leveling effect, the acid strength of binary compounds of hydrogen with nonmetals (A) increases as the H-A bond strength decreases down a group in the periodic table. For group 17, the order of increasing acidity is HF < HCl < HBr < HI. Likewise, for group 16, the order of increasing acid strength is H2O < H2S < H2Se < H2Te. Across a row in the periodic table, the acid strength of binary hydrogen compounds increases with increasing...