Optimal beam quality for chest flat panel detector system: realistic phantom study
Chie Kuwahara1, Takatoshi Aoki2, Nobuhiro Oda3
1Department of Radiology, University of Occupational and Environmental Health, 1-1 Iseigaoka, Yahatanishi-ku, Kitakyushu, 807-8555, Japan.
Objective:
To investigate optimal beam quality for chest flat panel detector (FPD) system by semi-quantitatively assessment using a realistic lung phantom.
Materials And Methods:
Chest FPD radiographs were obtained on a realistic lung phantom with simulated lung opacities using various X-ray tube voltage levels (90-140 kV) with/without copper filter. Entrance skin dose was set to maintain identical for all images (0.1 mGy). Three chest radiologists unaware of the exposure settings independently evaluated the image quality of each simulated opacity and normal structure using a 5-point scale (+ 2: clearly superior to the standard; + 1: slightly superior to the standard; 0: equal to the standard; - 1: slightly inferior to the standard; - 2: clearly inferior to the standard). The traditional FPD image obtained at a tube voltage of 120 kV was used as the standard. The scores of image quality were statistically compared using the Wilcoxon rank test with Bonferroni correction.
Results:
FPD images using 90-kV shot with copper filter were superior to the traditional 120-kV shot without filter with respect to the visibility of vertebra, pulmonary vessels, and nodules overlapping diaphragm and heart (p < 0.05). There was no significant difference with respect to the visibility of all other simulated lung opacities (lung nodules except for overlying diaphragm/heart and honeycomb opacity) between each tube voltage level with/without copper filter and the traditional 120-kV shot without filter.
Conclusion:
Image quality of FPD images using 90 kV with copper filtration is superior to that using standard tube voltage when dose is identical.
Key Points:
• FPD image quality using 90 kV with filter is superior to that using traditional beam. • Ninety-kilovolt shot with copper filter may be suitable for chest FPD image. • Clinical study dealing with chest FPD beam optimization would be warranted.
Related Concept Videos
Wood Panel Products
Flail Chest-I
Flail chest is a severe and potentially life-threatening condition characterized by the fracture of three or more adjacent ribs in multiple places. It is most commonly caused by direct impacts and trauma, such as motor vehicle accidents or injuries from a steering wheel impact. It can also occur due to falls in elderly individuals with osteoporosis, or assaults involving sharp objects.
Pathophysiology
The pathophysiology of flail chest is complex, involving fractures of...
Flail Chest-II
Assessment:
1. Clinical Evaluation:
History:
Chest Physiotherapy
Purpose
CPT is primarily used for patients with excessive bronchial secretions who have difficulty clearing...
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...