Author Spotlight: Optimized Lung MRI Protocol with Computationally Efficient Reconstruction Methods
Published on: September 6, 2024
A black-blood ultra-short echo time (UTE) sequence for 3D isotropic resolution imaging of the lungs
Jean Delacoste1, Helene Feliciano1, Jérôme Yerly1,2
1Department of Radiology, University Hospital (CHUV) and University of Lausanne (UNIL), Lausanne, Switzerland.
Purpose:
Ultra-short echo time MRI is a promising alternative to chest CT for cystic fibrosis patients. Black-blood imaging in particular could help discern small-sized anomalies, such as mucoid plugging, which may otherwise be confused with neighboring blood vessels, particularly when contrast agent is not used. We, therefore, implemented and tested an ultra-short echo time sequence with black-blood preparation. Additionally, this sequence may also be used to generate bright-blood angiograms.
Methods:
Using this sequence, data was acquired during free breathing in 10 healthy volunteers to obtain respiratory-motion-resolved 3D volumes covering the entire thorax with an isotropic resolution of (1 mm)3 . The magnitude of signal suppression relative to a bright-blood reference acquisition was quantified and compared with that obtained with a turbo-spin echo (TSE) acquisition. Bright-blood angiograms were also generated by subtraction. Finally, an initial feasibility assessment was performed in 2 cystic fibrosis patients, and images were visually compared with contrast-enhanced images and with CT data.
Results:
Black-blood preparation significantly decreased the average normalized signal intensity in the vessel lumen (-66%; P < 0.001). Similarly, blood signal was significantly lowered (-60%; P = 0.001) compared with the TSE acquisition. In patients, mucoid plugging could be emphasized in the black-blood datasets. An intercostal artery could also be visualized in the subtraction angiograms.
Conclusion:
Black-blood free-breathing ultra-short echo time imaging was successfully implemented and motion-resolved full volumetric coverage of the lungs with high spatial resolution was achieved, while obtaining an angiogram without contrast agent injection. Encouraging initial results in patients prompt further investigations in a larger cohort.
Related Concept Videos
Echo
Imagine the sound is reflected back to the ears. Assuming that the source is very close to the human, the difference between hearing the two sounds—the emitted sound and the reflected sound—may be more than the minimum time for perceiving distinct sounds. If this is the case,...
Lung Capacity
Detection of Black Holes
Their closest cousins are neutron stars, which are composed almost entirely of neutrons packed against each other, making them extremely dense. A neutron star has the same mass as the Sun but its diameter is only a few kilometers. Therefore, the escape velocity from their surface is close to the speed of light.
Not until the 1960s, when the first neutron...
Cis-regulatory Sequences
Short-distance Transport of Resources
Power System Three-Phase Short Circuits