Treatment response assessment in [18F]FDG-PET/CT oncology scans: Impact of count statistics variation and
Carlotta Dolci1, Chiara Spadavecchia1, Cinzia Crivellaro2
1University of Milan-Bicocca, School of Medicine and Surgery, Milan, Italy.
Purpose:
To investigate influences of reconstruction algorithms and count statistics variation on quantification and treatment response assessment in cancer patients, by using a large field of view-FOV scanner.
Methods:
54 cancer patients underwent PET/CT scan: 1) at baseline: 1.5 min/FOV, reconstructed by ordered-subset expectation maximization + point-spread-function-OSEM-PSF and bayesian penalised-likelihood-BPL algorithm 2) at restaging: 2 min/FOV, reconstructed also at 1.5 and 1 min/FOV, using OSEM-PSF and BPL. SUL (lean-body mass SUV) peak and max were measured for each target-lesion (n = 59). Differences in quantification obtained from datasets with different reconstruction algorithms and different time/FOV were evaluated. For any pair of PET datasets, metabolic response was assessed by using SULpeak, with a threshold of 30% in variation considered as significant.
Results:
Both at baseline and restaging, SULpeak and max values were higher in BPL reconstructions than in OSEM-PSF (p < 0.0001). SULpeak at different time/FOV reconstructions showed no statistically significant differences both with OSEM-PSF and BPL; SULmax depended on acquisition time (p < 0.05). In 56/59 lesions (95%) therapy response was concordant regardless count statistics variation and reconstruction algorithm; 2/59 (3%) showed different responses according to count statistics, both for OSEM-PSF and BPL; in 1/59 lesion (2%) response was different depending on reconstruction algorithm used.
Conclusions:
BPL provided higher SULpeak and max than OSEM-PSF. With a large FOV/high sensitivity scanner, variation of time/FOV in restaging PET scans gave stable and reproducible results in terms of SULpeak, both for OSEM-PSF and BPL. Thus, metabolic response defined by SULpeak variation proved to be quite independent from count statistics.
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Statistical Significance
What is Variation?
The range, standard deviation, standard error, and variance are the different measures of variation.
Range: The range is the difference between its maximum and...
Variation
When independent and dependent variables are plotted on a scatter plot, the slope of a line is a value that describes the rate of change between the two...
Probability in Statistics
An example of a simple event is a coin toss. The result of a coin toss is either a head or a tail. Here, head and tail are two simple events. These two simple events make up the sample space. Further, the probability of an event occurring falls within the range of 0 to 1. The probability of an...
Leaky Scanning