Evaluation of deformable image registration accuracy for CT images of the thorax region
Sebastian Sarudis1, Anna Karlsson2, Dan Bibac3
1Department of Radiation Physics, Sahlgrenska Academy, University of Gothenburg, 413 90 Göteborg, Sweden; Department of Medical Physics, County Hospital Ryhov, 551 85 Jönköping, Sweden.
Purpose:
Evaluate the performance of three commercial deformable image registration (DIR) solutions on computed tomography (CT) image-series of the thorax.
Methods:
DIRs were performed on CT image-series of a thorax phantom with tumor inserts and on six 4-dimensional patient CT image-series of the thorax. The center of mass shift (CMS), dice similarity coefficient (DSC) and dose-volume-histogram (DVH) parameters were used to evaluate the accuracy. Dose calculations on deformed patient images were compared to calculations on un-deformed images for the gross tumor volume (GTV) (Dmean, D98%), lung (V20Gy, V12Gy), heart and spinal cord (D2%).
Results:
Phantom structures with constant volume and shifts ≤30 mm were reproduced with visually acceptable accuracy (DSC ≥ 0.91, CMS ≤ 0.9 mm) for all software solutions. Deformations including volume changes were less accurate with 9/12 DIRs considered visually unacceptable. In patients, organs were reproduced with DSC ≥ 0.83. GTV shifts ≤1.6 cm were reproduced with visually acceptable accuracy by all software while larger shifts resulted in failures for at least one of the software. In total, the best software succeeded in 18/25 DIRs while the worst succeeded in 12/25 DIRs. Visually acceptable DIRs resulted in deviations ≤3.0% of the prescribed dose and ≤3.6% of the total structure volume in the evaluated DVH-parameters.
Conclusions:
The take home message from the results of this study is the importance to have a visually acceptable registration. DSC and CMS are not predictive of the associated dose deviation. Visually acceptable DIRs implied dose deviations ≤3.0%.
More Related Videos
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Muscles of the Thorax
The diaphragm is at the core of thoracic musculature, the primary muscle involved in breathing. This expansive, dome-shaped muscle marks the division between the thoracic and abdominal cavities. It...
Veins of Thorax
The azygos vein, positioned just right of the midline and anterior to the vertebral column, begins at the junction of the right ascending lumbar and subcostal veins, terminating in the superior vena cava. This vein drains blood from the right side of the thoracic wall, thoracic viscera, and posterior abdominal wall.
The...
Improving Translational Accuracy