Evaluation of the detection of blasts by Sysmex hematology instruments, CellaVision DM96, and manual microscopy using
Heidi Eilertsen1,2, Per Christian Saether1, Carola E Henriksson3,4
1Department of multidisciplinary laboratory medicine and medical biochemistry, Akershus University Hospital, Lørenskog, Norway.
Introduction:
The aim of the present study was to evaluate the diagnostic ability of blast flags generated by Sysmex instruments (XE/XN) by comparing with immunophenotyping by flow cytometry (IFCM). Additionally, the ability of manual microscopy and CellaVision DM96 (pre- and reclassification) to predict the presence of "true" blasts was investigated.
Methods:
Blood samples (n = 240) with suspect pathology flags reported by the XE were collected from the daily workload and examined by the XN, by manual microscopy, by CellaVision DM96 and by IFCM (CytoDiff Panel).
Results:
The ROC analysis for blasts showed an area under the curve of 0.64 ("Blasts?") (XE), 0.57 ("Blasts/Abn Lympho?") (XN), 0.75 (CellaVision preclassification procedure), 0.78 (CellaVision reclassification procedure), and 0.81 (manual microscopy). The sensitivity of blast detection varied between the methods from 0.41 (XE) to 0.90 (XN), and the specificity varied from 0.17 (XN) to 0.95 (CellaVision reclassification).
Conclusions:
The CellaVision reclassification procedure has a diagnostic ability for predicting blasts close to that of manual microscopy. The blood smear methods show a notable number of false negative results. The Sysmex XN reported a higher rate of true positive blast flags than the XE. Taken together, the CytoDiff method could be a useful alternative to smear examination to correctly identify blasts.
Related Concept Videos
Flow Cytometry
In...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
Instrument Transformers
Uncertainty in Measurement: Reading Instruments
Capillary Electrophoresis: Instrumentation