Related Experiment Video
Updated: Jan 29, 2026

Author Spotlight: Advancements in 3D Optical Imaging for Comprehensive Body Composition Assessment in Modern Research
Published on: June 7, 2024
Optical Topographic Imaging for Spinal Intraoperative Three-Dimensional Navigation in Mini-Open Approaches: A
Daipayan Guha1, Raphael Jakubovic2, Naif M Alotaibi3
1Division of Neurosurgery, Department of Surgery, University of Toronto. Toronto, Toronto, Ontario, Canada; Institute of Medical Science, School of Graduate Studies, University of Toronto. Toronto, Toronto, Ontario, Canada; Biophotonics and Bioengineering Laboratory, Sunnybrook Health Sciences Centre. Toronto, Toronto, Ontario, Canada.
Objective:
Computer-assisted three-dimensional navigation often guides spinal instrumentation. Optical topographic imaging (OTI) offers comparable accuracy and significantly faster registration relative to current navigation systems in open posterior thoracolumbar exposures. We validate the usefulness and accuracy of OTI in minimally invasive spinal approaches.
Methods:
Mini-open midline posterior exposures were performed in 4 human cadavers. Square exposures of 25, 30, 35, and 40 mm were registered to preoperative computed tomography imaging. Screw tracts were fashioned using a tracked awl and probe with instrumentation placed. Navigation data were compared with screw positions on postoperative computed tomography imaging, and absolute translational and angular deviations were computed. In vivo validation was performed in 8 patients, with mini-open thoracolumbar exposures and percutaneous placement of navigated instrumentation. Navigated instrumentation was performed in the previously described manner.
Results:
For 37 cadaveric screws, absolute translational errors were (1.79 ± 1.43 mm) and (1.81 ± 1.51 mm) in the axial and sagittal planes, respectively. Absolute angular deviations were (3.81 ± 2.91°) and (3.45 ± 2.82°), respectively (mean ± standard deviation). The number of surface points registered by the navigation system, but not exposure size, correlated positively with the likelihood of successful registration (odds ratio, 1.02; 95% confidence interval, 1.009-1.024; P < 0.001). Fifty-five in vivo thoracolumbar pedicle screws were analyzed. Overall (mean ± standard deviation) axial and sagittal translational errors were (1.79 ± 1.41 mm) and (2.68 ± 2.26 mm), respectively. Axial and sagittal angular errors were (3.63° ± 2.92°) and (4.65° ± 3.36°), respectively. There were no radiographic breaches >2 mm or any neurovascular complications.
Conclusions:
OTI is a novel navigation technique previously validated for open posterior exposures and in this study has comparable accuracy for mini-open minimally invasive surgery exposures. The likelihood of successful registration is affected more by the geometry of the exposure than by its size.
Related Concept Videos
Local Anesthetics: Clinical Application as Spinal Anesthesia
Initiation of Translation
First, the initiator tRNA must be selected from the pool of elongator tRNAs by eukaryotic initiation factor 2 (eIF2). The initiator tRNA (Met-tRNAi) has conserved sequence elements including modified bases at...
Initiation of Translation
Topographic Surveying and Contours
Plotting of Topographic Maps
Aneurysm II: Clinical Manifestations and Diagnostic Studies