Related Experiment Video
Updated: Jan 29, 2026

Design and Development of Aptamer–Gold Nanoparticle Based Colorimetric Assays for In-the-field Applications
Published on: June 23, 2016
Portable Smartphone-based Colorimetric Analyzer with Enhanced Gold Nanoparticles for On-site Tests of Seafood Safety
Longjie Zhong1,2, Jiadi Sun1, Ying Gan1
1Key Laboratory for Biomedical Engineering of Ministry of Education, Biosensor National Special Laboratory, Department of Biomedical Engineering, Zhejiang University.
Abstract:
Saxitoxin (STX) is one of the paralytic shellfish poisons (PSP) that endanger people's health. It is necessary to develop methods for the on-site rapid detection for STX in order to prevent safety accidents. An enzyme-linked immunosorbent assay (ELISA) is timesaving and effective, but it is not suitable for large-scale in-field tests due to the expensiveness of commercial ELISA kits and the bulkiness of a microtiter plate reader (MTPR). In this study, a portable smartphone-based colorimetric analyzer (SBCA) with a cost-effictive enhanced gold nanoparticle-based ELISA (EGNB-ELISA) was proposed for STX detection. In a bicinchoninic acid (BCA) protein assay (R2 = 0.9939) and a glucose assay (R2 = 0.9937), SBCA was shown to be in good agreement with MTPR. EGNB-ELISA had a 12.5-fold lower detection limit (0.4 ng/mL) and a lower detection range (1 - 50 ng/mL, Y = 0.4037X + 0.3564, R2 = 0.9797) than the classical ELISA. The recovery rate ranged over 89.1 - 112.2%. The whole detection system, combining both homemade SBCA and ENGB-ELISA, is expected to satisfy the needs of on-site STX sample tests to guarantee seafood safety.
Related Concept Videos
Survey Safety
Household Wiring And Electrical Safety
Self-Evaluation: Self-Enhancement and Self-Verification
Bioavailability Enhancement: Drug Solubility Enhancement
Bioavailability Enhancement: Drug Permeability Enhancement
Bioavailability Enhancement: Drug Stability Enhancement and GI Retention